8.3 Cleaning the wet measuring cell
The measuring cell glasses do not need to be removed
for cleaning. A disassembly is only necessary for an
exchange or if there is heavy soiling.
‘Removing the measuring cell glasses’ on page 45
Use the cloths included in the service kit for cleaning the measuring cell
glasses. Alcohol can be used for stubborn soiling.
8.3.1 Emptying the system
1.
Check whether the hose at the outlet connector for the measuring
liquid is connected and ends in a collecting vessel or a drain.
2.
Start the MaScontrol software.
3.
Create a new root folder or use an already existing one in order to
open or create a folder.
4.
Double-click the symbol in front of the newly created folder. A
measuring results folder opens.
5.
From this measuring results folder you can select the SOP "Drain"
by clicking on "Measurement".
SOP refers to a Standard Operating Procedure
used in our software. In an SOP, work steps are
pre-defined and processed with a single key-
stroke.
Ensure that you select the SOP which is suitable
for your device.
6.
Click on "Start Measurement" to empty the system.
Cleaning
- 38 -