Chapter 4: Operation
Fresenius Medical Care multiFiltrate IFU-EN-UK 15A-2015
4-119
System parameters
4.12 System parameters
Key sound
The audible feedback indicating that a key was pressed can be turned
on or off.
Audible alarm volume
The volume of audible signals can be adjusted in this field.
Display brightness
The brightness of the screen can be adjusted from 50–100% in
increments of 5%.
Functional test sequence
When
Detailed
is selected, all test steps will be displayed on the
screen. When
Automatic
is selected, only the display test will be
shown.
Version
This field can be used to view the software version.
Line system
If "Standard" or "Cassette" is selected, then the screen will support the
insertion of the line system as appropriate. If a treatment mode has
been chosen where no selection is possible, then the menu field is
greyed out.
Tubing arrangement
When
Detailed
is selected, the installation of the line system will be
shown step by step. When
Complete
is selected, the fully inserted line
system will be displayed.
Note
The responsible organisation should define the most important
configurable parameter settings itself, or confirm the default values and
have these set by service support as required.
The
System parameters
menu can be
accessed at any time. The fields
Key sound
,
Audible alarm volume
, and
Display
brightness
can always be changed. All other
parameter fields and submenus are available
only until
Start priming
is executed for the
tubing system or at the end of the treatment.
For Ci-Ca treatments, the
Cassette for
adults
is selected under
Line system
and
cannot be changed. The field
mDL settings
is only available when an mDL is integrated
and activated.
The selected field must be confirmed by
pressing the rotary selector
[OK]
.
Summary of Contents for multiFiltrate
Page 1: ...multiFiltrate Instructions for Use Software version 5 3 Edition 15A 2015 Part no M56 963 1 ...
Page 2: ......
Page 12: ...xii Fresenius Medical Care multiFiltrate IFU EN UK 15A 2015 ...
Page 16: ...Chapter 1 Index 1 4 Fresenius Medical Care multiFiltrate IFU EN UK 15A 2015 ...
Page 42: ...Chapter 3 Design 3 10 Fresenius Medical Care multiFiltrate IFU EN UK 15A 2015 ...
Page 186: ...Chapter 5 Alarm processing 5 20 Fresenius Medical Care multiFiltrate IFU EN UK 15A 2015 ...
Page 188: ...Chapter 6 Cleaning disinfection 6 2 Fresenius Medical Care multiFiltrate IFU EN UK 15A 2015 ...
Page 218: ...Chapter 7 Functional description 7 30 Fresenius Medical Care multiFiltrate IFU EN UK 15A 2015 ...
Page 228: ...Chapter 9 Installation 9 4 Fresenius Medical Care multiFiltrate IFU EN UK 15A 2015 ...
Page 252: ...Chapter 12 Specifications 12 18 Fresenius Medical Care multiFiltrate IFU EN UK 15A 2015 ...
Page 260: ...Chapter 13 Definitions 13 8 Fresenius Medical Care multiFiltrate IFU EN UK 15A 2015 ...
Page 266: ...Chapter 14 Options 14 6 Fresenius Medical Care multiFiltrate IFU EN UK 15A 2015 ...
Page 282: ...Chapter 15 Appendix 15 16 Fresenius Medical Care multiFiltrate IFU EN UK 15A 2015 ...