Chapter 4: Operation
72
Fresenius Medical Care
AquaA
IFU-EN-US 06B-2021
4.10.2 STATUS–Report
Menu structure–overview
4.10.2.1 Daily report AquaA
The daily reports are displayed by pressing
the
Report
button.
In the daily report, the current data of the
device in the
SUPPLY
mode are recorded
every day at a programmed time. These data
are intended to assist the technician when
analyzing the device.
The following data are recorded:
– Date and time when the data was
collected
– Feed conductivity
CD-F
– Feed temperature
T-F
– Dialysis water conductivity
CD-P
– Dialysis water temperature
T-P
– Concentrate pressure
P-C
– Permeate pressure
P-P
– Feed flow
FL-F
– Maximum feed flow
FL-Fmax
Tip
More values are shown on the second page.
Report
Page 1
Page 2