Chapter 4: Operation
Fresenius Medical Care
AquaA
IFU-EN-US 06B-2021
51
4.5.4 SUPPLY–Permeate stop
Dialysis water conductivity and temperature monitoring
If the conductivity or temperature alarm limit is exceeded, the dialysis
water supply will be stopped by the closing of the dialysis water flow
valve. Yield control is deactivated during this time.
4.5.5 SUPPLY–Warning
During
SUPPLY–Warning
mode, the
AquaA
is still operational but
requires an analysis of the warning (see chapter 5.3.1 on page 111).
The display shows the current values or the list of messages with the
current message.
4.5.6 SUPPLY–External locking
In this operating mode, the functionality of the
AquaA
is limited due to
a water pretreatment system signal. There is an insufficient supply of
water to the
AquaA
. The supply of water from the water pretreatment
system to the
AquaA
is blocked as a precautionary measure. A run-dry
protection warning will therefore appear if dialysis water is consumed.
Once the water pretreatment system reports an adequate water supply,
the valve will be reopened.
The external locking function is configured by the service technician in
the Service menu of the
AquaA
.
Summary of Contents for AquaA
Page 2: ......
Page 10: ...10 Fresenius Medical Care AquaA IFU EN US 06B 2021 ...
Page 32: ...Chapter 2 Important information 32 Fresenius Medical Care AquaA IFU EN US 06B 2021 ...
Page 108: ...Chapter 4 Operation 108 Fresenius Medical Care AquaA IFU EN US 06B 2021 ...
Page 152: ...Chapter 9 Installation 152 Fresenius Medical Care AquaA IFU EN US 06B 2021 ...
Page 186: ...Chapter 13 Definitions 186 Fresenius Medical Care AquaA IFU EN US 06B 2021 ...
Page 217: ...Chapter 14 Options Fresenius Medical Care AquaA IFU EN US 06B 2021 217 ...
Page 218: ...Chapter 14 Options 218 Fresenius Medical Care AquaA IFU EN US 06B 2021 ...