![Fresenius Medical Care AquaA Instructions For Use Manual Download Page 239](http://html.mh-extra.com/html/fresenius-medical-care/aquaa/aquaa_instructions-for-use-manual_2331674239.webp)
Chapter 15: Appendix
Fresenius Medical Care
AquaA
IFU-EN-US 06B-2021
239
15.7 Collecting a sample for chemical analysis
15.7.1 Preparation
Dialysis water sampling is only possible if the reverse osmosis system
is in the
SUPPLY
mode, or if it is producing dialysis water during a
manual rinse program in the
RINSE
mode.
Before collecting the sample, the reverse osmosis system must have
been in operation for at least 20 minutes. If the device is not in the
SUPPLY
mode, the manual rinse program must be started.
The sample is collected in the
SUPPLY
or
RINSE
mode.
15.7.2 Accessories, equipment
The following equipment is recommended by the manufacturer:
– Rubber gloves
For the chemical sampling, use the sample containers provided by the
laboratory. The bag with adapter (part number: G03000832) can be
used as dialysis water sampling equipment.
15.7.3 Collecting a sample for chemical analysis
Summary of Contents for AquaA
Page 2: ......
Page 10: ...10 Fresenius Medical Care AquaA IFU EN US 06B 2021 ...
Page 32: ...Chapter 2 Important information 32 Fresenius Medical Care AquaA IFU EN US 06B 2021 ...
Page 108: ...Chapter 4 Operation 108 Fresenius Medical Care AquaA IFU EN US 06B 2021 ...
Page 152: ...Chapter 9 Installation 152 Fresenius Medical Care AquaA IFU EN US 06B 2021 ...
Page 186: ...Chapter 13 Definitions 186 Fresenius Medical Care AquaA IFU EN US 06B 2021 ...
Page 217: ...Chapter 14 Options Fresenius Medical Care AquaA IFU EN US 06B 2021 217 ...
Page 218: ...Chapter 14 Options 218 Fresenius Medical Care AquaA IFU EN US 06B 2021 ...