5.4 Error category 01–system and hardware problems
The following tables list all errors which might occur when operating the
system.
As some messages are identical except for certain criteria, these have
been combined into groups. The errors are classified in categories of
01 to 03.
An error code marked with „*“ denotes a self-confirming message.
Error code
Visual
indica-
tor
Messages
Cause
Action required
F–01–01–01
Signal:
red
FAILURE:
Replace terminal
battery
– Insufficient display battery
capacity
Contact the service
department
F–01–01–02
Signal:
red
FAILURE:
Overvoltage
– Excess voltage at the
operating unit.
– This message is
automatically cleared once
the specified supply voltage
has been reached.
Check display power
supply
Contact the service
department
F–01–01–03
Signal:
red
FAILURE:
FATAL ERROR
Code: , Subcode:
– This message is generated
by the operating system of
the terminal if proper
operation can not be
continued due to lack of
security.
To reproduce the
problem which occurred,
the code and subcode as
well as the software
versions of the operating
system and the user
interface must be known.
Contact the service
department
F-01–01–04
Signal:
red
FAILURE:
COMMUNICATION
ERROR Code:,
Subcode:
– Protocol and interface error
To reproduce the
problem which occurred,
the code and subcode as
well as the software
versions of the operating
system and the user
interface must be known.
Contact the service
department
F–01–01–05
Signal:
red
FAILURE:
I/O-Bus
– BUS system connection
interrupted
– BUS component defective
Contact the service
department
Summary of Contents for AquaA
Page 2: ......
Page 10: ...10 Fresenius Medical Care AquaA IFU EN US 06B 2021 ...
Page 32: ...Chapter 2 Important information 32 Fresenius Medical Care AquaA IFU EN US 06B 2021 ...
Page 108: ...Chapter 4 Operation 108 Fresenius Medical Care AquaA IFU EN US 06B 2021 ...
Page 152: ...Chapter 9 Installation 152 Fresenius Medical Care AquaA IFU EN US 06B 2021 ...
Page 186: ...Chapter 13 Definitions 186 Fresenius Medical Care AquaA IFU EN US 06B 2021 ...
Page 217: ...Chapter 14 Options Fresenius Medical Care AquaA IFU EN US 06B 2021 217 ...
Page 218: ...Chapter 14 Options 218 Fresenius Medical Care AquaA IFU EN US 06B 2021 ...