Chapter 11: System Description
Fresenius Medical Care 5008 OP 5/09.06
11-27
Conductivity and temperature sensors check the dialysate for the
correct mixing ratio and temperature. If these are not within the set
alarm limits, the dialyzer supply valve closes and the bypass valve
opens, therefore discharging any incorrect dialysate directly to the
drain.
Transmembrane pressure is calculated from the values of the dialysate
pressure transducers. The blood leak detector generates an alarm if
any blood loss is detected in the dialysate. The preset correct fluid
removal from the patient is achieved by the UF pump whilst the flow
pump controls the dialysate flow from the dialyzer to the balancing
chamber. The used dialysate is then passed via the heat exchanger to
the drain to be removed.
When the disinfection programs are in progress, disinfectant is supplied
via the mixing system and the hydraulics to clean and disinfect the unit.
11.5.4 Description bi
b
ag
®
When using the bi
b
ag
®
, a bag containing sodium bicarbonate powder
(NaHCO
3
) is used instead of a canister containing a liquid bicarbonate
solution.
The powder is prepared in the hemodialysis system to produce ready-
to-use bicarbonate.
For recommended values for the maximum treatment times in relation
to the package size used and the dialysate flow refer to the table below.
11.5.5 OCM Description
O
Information regarding OCM
OCM logo
OCM
®
is a registered trademark.
Intended use
Fields of application
The OCM allows determination of the average effective urea clearance
(K), the dialysis dose Kt/V and the plasma sodium concentration during
dialysis.
Package size
Maximum treatment times
bi
b
ag
®
650 g (NaHCO
3
)
300 ml/min approx. 10 hours
500 ml/min approx. 6 hours
800 ml/min approx. 4 hours
bi
b
ag
®
900 g (NaHCO
3
)
300 ml/min approx. 15 hours
500 ml/min approx. 9 hours
800 ml/min approx. 6 hours
Summary of Contents for 5008
Page 2: ......
Page 14: ...Chapter 1 Index 1 6 Fresenius Medical Care 5008 OP 5 09 06 ...
Page 38: ...Chapter 3 Design 3 16 Fresenius Medical Care 5008 OP 5 09 06 ...
Page 48: ...Chapter 4 Graphical User Interface 4 10 Fresenius Medical Care 5008 OP 5 09 06 ...
Page 49: ...Fold Out Sheet 5 Preparation Fresenius Medical Care 5008 OP 5 09 06 ...
Page 66: ...Chapter 5 Preparation 5 16 Fresenius Medical Care 5008 OP 5 09 06 ...
Page 96: ...Chapter 5 Preparation 5 46 Fresenius Medical Care 5008 OP 5 09 06 ...
Page 111: ...Fold Out Sheet 6 Treatment Fresenius Medical Care 5008 OP 5 09 06 ...
Page 112: ...TREATMENT SCREEN SN setting the level in the SN chamber ...
Page 120: ...Chapter 6 Treatment 6 8 Fresenius Medical Care 5008 OP 5 09 06 ...
Page 128: ...Chapter 6 Treatment 6 16 Fresenius Medical Care 5008 OP 5 09 06 ...
Page 135: ...Fold Out Sheet 7 Reinfusion Fresenius Medical Care 5008 OP 5 09 06 ...
Page 136: ...REINFUSION SCREEN ...
Page 140: ...Chapter 7 Reinfusion 7 4 Fresenius Medical Care 5008 OP 5 09 06 ...
Page 144: ...Chapter 7 Reinfusion 7 8 Fresenius Medical Care 5008 OP 5 09 06 ...
Page 146: ...Chapter 7 Reinfusion 7 10 Fresenius Medical Care 5008 OP 5 09 06 ...
Page 147: ...Fold Out Sheet 8 Cleaning Fresenius Medical Care 5008 OP 5 09 06 ...
Page 148: ...Connecting the disinfectant container CLEANING SCREEN Program run graphic e g Diasteril ...
Page 155: ...Fold Out Sheet 9 Alarm processing Fresenius Medical Care 5008 OP 5 09 06 ...
Page 175: ...Fold Out Sheet 10 Other Functions Fresenius Medical Care 5008 OP 5 09 06 ...
Page 210: ...Chapter 10 Other Functions 10 34 Fresenius Medical Care 5008 OP 5 09 06 ...
Page 231: ...Chapter 11 System Description Fresenius Medical Care 5008 OP 5 09 06 11 21 ...
Page 272: ...Chapter 13 Certificates 13 2 Fresenius Medical Care 5008 OP 5 09 06 ...
Page 273: ...Chapter 13 Certificates Fresenius Medical Care 5008 OP 5 09 06 13 3 ...
Page 274: ...Chapter 13 Certificates 13 4 Fresenius Medical Care 5008 OP 5 09 06 ...
Page 276: ...Chapter 14 Appendix 14 2 Fresenius Medical Care 5008 OP 5 09 06 ...
Page 277: ...Fold Out Sheet 15 OCM Fresenius Medical Care 5008 OP 5 09 06 ...
Page 278: ...OCM SCREEN OCM diagram ...
Page 284: ...Chapter 15 Option OCM Online Clearance Monitoring 15 6 Fresenius Medical Care 5008 OP 5 09 06 ...
Page 285: ...Fold Out Sheet 16 ONLINEplus Fresenius Medical Care 5008 OP 5 09 06 ...
Page 286: ...ONLINE SCREEN ...
Page 287: ...Chapter 16 ONLINEplus Fresenius Medical Care 5008 OP 5 09 06 16 1 16 ONLINEplus ...
Page 290: ...Chapter 16 ONLINEplus 16 4 Fresenius Medical Care 5008 OP 5 09 06 ...
Page 291: ...Fold Out Sheet 17 Option SN Fresenius Medical Care 5008 OP 5 09 06 ...
Page 292: ...SINGLE NEEDLE SCREEN ...
Page 296: ...Chapter 17 Option SN Single Needle 17 4 Fresenius Medical Care 5008 OP 5 09 06 ...
Page 297: ...Fold Out Sheet 18 Option BPM Fresenius Medical Care 5008 OP 5 09 06 ...
Page 304: ...Chapter 18 Option BPM Blood Pressure Monitoring 18 6 Fresenius Medical Care 5008 OP 5 09 06 ...
Page 306: ...Chapter 19 Option CBPM 19 2 Fresenius Medical Care 5008 OP 5 09 06 ...
Page 307: ...Fold Out Sheet 20 BTM Option Fresenius Medical Care 5008 OP 5 09 06 ...
Page 313: ...Fold Out Sheet 21 BVM Option Fresenius Medical Care 5008 OP 5 09 06 ...
Page 326: ...Chapter 23 Options BLK WET 23 4 Fresenius Medical Care 5008 OP 5 09 06 ...