2008K/K2 Troubleshooting Guide
P/N 507298 Rev. H
Page 506
Yes
BOTH Pressure Tests pass! Proceed to
page 510
, procedure number
No Both tests DO NOT pass! Proceed to
page 440
SECTION 9- TMP (PRESSURE) PROBLEMS
to perform related pressure/leak checks.
UF- 3.0.0 INADEQUATE UF PUMP OUTPUT
a) Per the Figure right, trace the UF Pump’s cable
to distribution board position “P22, UF-P”.
b) ENSURE the connector is placed
properly AND the cable is not damaged.
c)
IMPORTANT, if NOT already, place
the machine into RINSE!
d) Ensure the external flow indicator’s ‘bob’ is moving up and down!
e) Per the Figure right, remove the tubing
from the UF Pump white INPUT (bottom)
nozzle. Is there flow through Filter #74
(Yes or No)?
Yes
Flow through the filter!
See procedure number
(page 507).
No
No flow! This may be the
problem! Replace UF Filter
#74.
Figure 87 – UF Pump Springs / Seals/ Diaphragm