2008K/K2 Troubleshooting Guide
P/N 507298 Rev. H
Page 449
TMP- 2.0.3 NO EXTERNAL PUMP LEAKS / ISOLATE FILLING PROGRAM (FILACT)
a) Return the concentrate connectors to acid and bicarb.
b) Return to Dialysis Program (“Select Program”
→ ‘Dialysis’ →
‘CONFIRM)!
c) Call debug screen 1. Is
FILACT
(middle column) = 1?
Yes
FILACT
= 1! Proceed to
page 452
, procedure number
No
FILACT
= 0! See parts a THROUGH d below:
a)
FIGURE BELOW
, inside the distribution, unplug the Air Sensor’s female Connector i.e.
4
th
CONNECTOR,
5
th
position from the left.
NOTE:
If the machine is CBE equipped the CBE
board plugs into the distribution board and Connector plugs into the CBE board which
positions it higher than the other connectors.
b)
Leave the Air Sensor’s Connector unplugged till instructed!
Parts c and d next page