132
va07.1_001g_Eng.fm
Cali
bra
tions
7.1.4 Etal 3, Air sensor
The air sensor calibration consists in calibrating the air detector board in several phases
A new administration set must be used for this procedure.
Etal 3
, press "enter".
The main menu "
Air sensor
" displays.
Perform a flow rate measurement according to
instructions described in paragraph "4.5 Flow rate check".
Press "OK" to launch the resonance search.
Wait the stop of resonance phase
Resonance frequency calibration phase with a tube and
water:
Water tube
: the air detector value in mV.
Frequency
: current frequency value in KHz.
Water tube: the air detector value in mV. The value
must lie between
600 < X < 2400mV.
Frequency: the resonance frequency for this phase.
Press "OK" to validate and move to the next phase.
Resonance frequency calibration phase without a tube.
Remove the pump tube and shut the door.
Without the tube: the air detector value in mV. The
value must be X < 300mV.
Frequency: the resonance frequency for this phase.
Press "OK" to validate and move to the next phase.
Calibration resonance frequency, with tube and air
Put a tube with air in place and close the door.
Air detector value in mV. The value must lie
between X < 300mV.
Frequency: resonance frequency for this phase.
Press "OK" to validate and move to the next phase.
Press "OK" again to confirm and save the calibration
values.
If the values are out of range, a message specifies the
non coformity: press "Exit".
Restart the calibration procedure.
Summary of Contents for Volumat Agilia
Page 1: ...0 9 C E S Volumat Agilia Technical Manual ...
Page 6: ...66 TM_Volumat_Agilia_Eng_TDM fm ...
Page 8: ...8 va01_001e_Eng fm Introduction 1 2 Operation diagram ...
Page 12: ...12 va01_001e_Eng fm Introduction ...
Page 26: ...26 va02_001e_Eng fm Description and operation ...
Page 30: ...30 va03 1_001e_Eng fm Description of the menus ...
Page 62: ...62 va04 4_001e_Eng fm Preventive maintenance ...
Page 70: ...70 va05 1_001e_Eng fm Troubleshooting ...
Page 76: ...76 va06 0_001g_Eng fm Intervention procedures ...
Page 82: ...82 va06 0_0011g_Eng fm Intervention procedures ...
Page 86: ...86 va06 0_0021g_Eng fm Intervention procedures ...
Page 94: ...94 va06 0_0040g_Eng fm Fiches d intervention ...
Page 104: ...104 va06 0_0060g_Eng fm Fiches d intervention ...
Page 110: ...110 va06 0_0070g_Eng fm Fiches d intervention ...
Page 134: ...134 va07 1_001g_Eng fm Calibrations ...
Page 136: ...136 va08 0_001d_Eng fm Spare parts catalogue ...