12.3 Maintenance requirements
WARNING
■
Perform preventive maintenance at least once every
2 years
. This includes
battery and membrane replacement. To avoid pumping performance
deterioration, it is important to follow maintenance requirements.
■
Preventive maintenance must be performed by qualified and trained
technical personnel in compliance with the technical manual and
procedures.
■
The qualified personnel must be informed if the device is dropped or if
any malfunction occurs. In this case, the device must not be used. Please
contact your biomedical department or Fresenius Kabi.
■
When replacing components, only use Fresenius Kabi spare parts.
■
When using the device on a patient, no maintenance action must be
performed.
Life cycle of Amika pump: 10 years provided that the maintenance is properly performed
as described above.
12.4 Service policy and rules
For further information concerning device servicing or use, please contact our sales
representative or our Customer service.
If the device must be sent for servicing, contact Fresenius Kabi to have packaging shipped
to your facility.
Clean and disinfect the device, because of potential harm or risks to staff health. Then pack
it in the provided packaging and ship to Fresenius Kabi.
INFORMATION
Fresenius Kabi is not liable for loss or damage to the device during transport.
62