8.2
Electromagnetic compatibility and interference
guidance
Warning
Use of the
Ambix activ
pump adjacent to or stacked with other
equipment should be avoided because it could result in improper
operation. If such use is necessary, this equipment and the other
equipment should be observed to verify that they are operating
normally.
Use of accessories, transducers and cables other than those
specified or provided by the manufacturer of the
Ambix
activ
pump could result in increased electromagnetic emissions or
decreased electromagnetic immunity of the pump and result in
improper operation.
Portable RF communication equipment (including peripherals
such as antenna cables, internal and external antennas) should
be used no closer than 10 cm for cell phones and 30 cm for other
equipment, to any part of the
Ambix activ
pump, including
cables specified by the manufacturer. Otherwise, degradation
of the essential performance of
Ambix activ
pump could result.
The user might be required to take mitigation measures, such as
relocating or re-orienting the equipment.
If the
Ambix
activ
pump is placed near RF communication equipment
such as cell phones, DECT phones or wireless access points, RFID
reader and tags, ... it is essential to observe a minimum distance
between the
Ambix
activ
pump and the equipment specified above. If
the
Ambix
activ
pump causes harmful interference or if it is itself
disrupted, the user is encouraged to try to correct the interference by
one of the following actions:
- Reorient or relocate the
Ambix
activ
pump, the patient, or the
disruptive equipment.
- Change the routing of cables.
- Connect the
Ambix
activ
pump mains plug to a protected, backed-up,
or filtered supply, or directly to a UPS (uninterruptible power supply)
circuit.
- Increase the separation between the
Ambix
activ
pump and the
disruptive equipment.
- Connect the
Ambix
activ
pump into an outlet on a different circuit from
that to which the patient or disruptive equipment is connected.
If the problem remains, the pump shall not be used in such environment.
For further information on EMC compliance, please refer to the
Ambix
activ
Technical Manual.