![Fresenius Kabi Ambix activ Instructions For Use Manual Download Page 18](http://html.mh-extra.com/html/fresenius-kabi/ambix-activ/ambix-activ_instructions-for-use-manual_2331619018.webp)
18
3.3
Loading the set (patient not connected)
3.3.1
Installing the set in the pump
1.
Open the door with the lever.
2.
Insert the
activ
clamp into the intended cavity of the pump. Position
and lock the green section of the set into the green slot in the pump.
3.
Place the upper ring of the blue kink protector in the blue coloured
notch. Take care that the pumping segment of the tubing is in front
of the pump mechanism.
4.
Close the pump door and push the lever down. The
activ
clamp will
open automatically. Flow will be prevented by the pumping
mechanism until administration is started.
5.
If applicable, open roller clamp after set installation, before starting
infusion.
Pump
mechanism
Lower tube guide
(Sensor area)