61
6.2.22
Checking the End of Infusion Alarm
14.
Press
C
to change the syringe.
15.
Press the arrow keys to select "BD Precise 50 mL".
16.
Press
OK
.
17.
Press
or .
18.
Press the arrow keys to select
.
19.
Select a Medium limit of 500 mmHg. (9.67 PSI or 0.67 bar).
20.
Select a flow rate of 120 mL/h.
21.
Start the infusion at 120 mL/h.
22.
Check for the absence of an occlusion alarm.
23.
Check that the infusion LEDs are flashing.
24.
Check that the occlusion alarm is triggered at an "x" pressure value of:
430 mmHg (8.31 PSI or 0.6 bar) < x < 570 mmHg (11.02 PSI or 0.8 bar).
25.
Open the syringe barrel clasp, push the disengagement lever down, and move the plunger driver to
the right.
26.
Remove the dynamometer.
1.
Power the pump off and on again to exit the Options menu and return to standard operation.
2.
Push the disengagement lever down and move the plunger to the right.
3.
Install a "BD Precise" 50 mL syringe.
4.
Secure the syringe with the syringe barrel clasp.
5.
Position the syringe piston at a remaining VTBI value of at least 10 mL.
6.
Push the disengagement lever down, and move the plunger driver gently to the left until it is in contact
with the syringe, with the antisiphon arm positioned on the syringe piston.
7.
Release the disengagement lever.
8.
Select a flow rate of 200 mL/h.
9.
Start the infusion and check that the end of infusion pre alarm is triggered.
10.
Wait for the End of infusion alarm.
11.
Run Test 6 Displacement.
12.
Note the Position value and check that Position = 17.7 mm ± 0.5 mm
(0.7 in ±0.02 in).
Summary of Contents for Agilia SP
Page 139: ...139 17 Spare Parts Catalog Refer to the spare parts catalog of the relevant pump ...
Page 143: ...143 VTBI Volume to Be Infused W Watts Term Description ...
Page 147: ...Local Contacts for Servicing Write your contacts here ...
Page 148: ...11378 2_tm_agilia_sp_eng ...
Page 305: ...Local Contacts for Servicing Write your contacts here ...