60
6.2.21
Checking the Pressure Limit Alarm
Figure 6.6:
Installation of dynamometer
Test Description
Run this test to check the triggering of an occlusion alarm when the pressure downstream the
pump exceeds the pressure limit stated in the Pressure management menu.
Required Tools and Equipment
1 Dynamometer
Prerequisites
The dynamometer’s calibration certificate is valid.
1.
Access the Basic Profile configuration. See section 5.1, page 32.
2.
Press the arrow keys to select
User
.
3.
Press
enter
.
4.
Press the arrow keys to select
User 4: Pressure
.
5.
Press
enter
.
6.
Select:
Mode
=
3 levels
High
=
900 mmHg
(17.4 PSI or 1.20 bar)
Medium
=
500 mmHg
(9.67 PSI or 0.67 bar)
Low
=
200 mmHg
(3.87 PSI or 0.27 bar)
7.
Power the pump off and on again to exit the
Options
menu and return to standard operation.
8.
Open the syringe barrel clasp, push the disengagement lever down, and move the plunger driver to
the right.
9.
Place the dynamometer in the syringe cradle, with the flanges correctly inserted in the provided slot.
10.
Secure the dynamometer with the syringe barrel clasp.
11.
Slide the arm spacer to the left.
12.
Push the disengagement lever, and move the plunger driver gently to the left until it is in contact with
the plunger head of the dynamometer.
13.
Release the disengagement lever.
The arms must not be in contact with the arm spacer.
Summary of Contents for Agilia SP
Page 139: ...139 17 Spare Parts Catalog Refer to the spare parts catalog of the relevant pump ...
Page 143: ...143 VTBI Volume to Be Infused W Watts Term Description ...
Page 147: ...Local Contacts for Servicing Write your contacts here ...
Page 148: ...11378 2_tm_agilia_sp_eng ...
Page 305: ...Local Contacts for Servicing Write your contacts here ...