Servicing
24.1
Information on Device Servicing
24.2
Maintenance Requirements
If the device must be sent for servicing, proceed as follows:
1.
Contact
Fresenius Kabi
to have packaging shipped to your facility.
2.
Clean and disinfect the device.
3.
Pack the device in the provided packaging.
4.
Ship the device to
Fresenius Kabi
.
Information
Fresenius Kabi
is not liable for loss or damage to the device
during transport.
For more information on servicing, contact your
Fresenius Kabi
sales representative.
Warning
Perform preventive maintenance at least once every 3 years. This
includes replacing the battery.
To ensure the device continues to operate normally, follow the
instructions below:
Preventive maintenance should be performed by trained and
qualified technical personnel in compliance with the technical
manual and procedures. Only authorized service personnel
should attempt to repair the device.
The qualified personnel must be informed if the device is
dropped or if any malfunctions occur. In this case, do not use the
device and contact your biomedical department or
Fresenius Kabi
.
Failure to comply with these maintenance procedures could
damage the device and lead to a functional failure. Internal
inspection of the device requires compliance with special
procedures to avoid damage to the device.
When replacing components, only use spare parts from
Fresenius Kabi
.
Summary of Contents for Agilia SP
Page 139: ...139 17 Spare Parts Catalog Refer to the spare parts catalog of the relevant pump ...
Page 143: ...143 VTBI Volume to Be Infused W Watts Term Description ...
Page 147: ...Local Contacts for Servicing Write your contacts here ...
Page 148: ...11378 2_tm_agilia_sp_eng ...
Page 305: ...Local Contacts for Servicing Write your contacts here ...