109
14.3
Preparing the Device for Storage
14.4
Using the Device After Storage
Prepare the device for storage as follows:
1.
Power the pump OFF and remove the disposable.
2.
If necessary (long-term storage), disconnect the pump’s power cord
and all data communication cables.
3.
Remove the pump from its mounting point.
4.
Clean the pump.
5.
Handle the pump with care, and store it in a compliant area.
For detailed instructions, refer to the related chapters in this document.
The device can be used immediately after storage without any cooling
or warm up period.
If the battery has been removed for long-term storage, contact your
biomedical department in order to reinstall the battery prior to use.
We recommend charging the battery for at least 6 hours.
We recommend that the "User test" is performed when the device is
installed after storage, and before being used on a patient, see section