97
12.3
General Remarks
12.4
List of Alarms
12.4.1
Syringe Alarms
When two alarms occur at the same time, the higher priority
alarm is displayed.
When two alarms with the same priority level are triggered at the
same time, the pump software assigns them a priority.
When the cause of a high-priority alarm has been fixed, the red
indicators switch off. However, the message remains displayed
at the top of the screen as a reminder of the cause of the alarm.
The device guarantees the triggering of high-level priority alarms
in every use condition.
A maximum of 1 mL may be infused due to a single fault
condition.
For all alarms (except occlusion alarms), the amount of time
between the alarm condition and the alarm generation is less
than 5 seconds.
If the AC power is disconnected and if the battery is discharged,
the alarms settings are not modified and are stored indefinitely.
Message
Priority
Stops
Infusion?
Problem / Resolution
Syringe
installation !!!
High (!!!)
Yes
The syringe is not installed correctly (plunger
driver, syringe barrel clasp or flange
detection).
Check the syringe installation.
Plunger head
alarm !!!
High (!!!)
Yes
The plunger head is missing or incorrectly
inserted.
Check the syringe installation.
Disengagement
mechanism !!!
High (!!!)
Yes
Disengaged mechanism.
Check the syringe installation.
Remove completely
syringe !
Low (!)
No
Preventive auto-test on potential failure of
plunger head.
Remove and reinstall the syringe.
Summary of Contents for Agilia SP
Page 139: ...139 17 Spare Parts Catalog Refer to the spare parts catalog of the relevant pump ...
Page 143: ...143 VTBI Volume to Be Infused W Watts Term Description ...
Page 147: ...Local Contacts for Servicing Write your contacts here ...
Page 148: ...11378 2_tm_agilia_sp_eng ...
Page 305: ...Local Contacts for Servicing Write your contacts here ...