57
7.10
Completing an Infusion
7.10.1
Near End of Infusion Alert
Prior to the end of an infusion, a
near end of infusion
alert is
automatically triggered. The following happens:
An audible alarm is triggered.
An alarm message appears on the pump screen.
The infusion indicator lights flash yellow.
Near end of infusion alert is triggered when the two criteria below are
reached simultaneously.
Setting
Range of Values
Default Pump
Setting
Time Before the End of the Infusion
From 1 to 30 minutes
5 minutes
The remaining volume of fluid in the
syringe < 10 % of the syringe capacity
N/A
N/A
Near end of infusion alert settings are configurable with Agilia
Vigilant Drug’Lib software (custom profiles), or in the pump options
(Basic Profile). For more information, refer to the technical manual.
Silencing Near End of Infusion Alert
1.
Press
to silence the alarm.
2.
If required, press
OK
to confirm the
empty syringe
mode.
Depending on the pump configuration, the following happens:
The infusion continues at the programmed rate until the plunger
reaches the tip of the syringe. The syringe is not completely
emptied (regular end of infusion).
The infusion continues until the syringe is completely emptied.
The flow rate decreases when the plunger reaches the tip of the
syringe (empty syringe - only in Simple Rate mode).
Summary of Contents for Agilia SP
Page 139: ...139 17 Spare Parts Catalog Refer to the spare parts catalog of the relevant pump ...
Page 143: ...143 VTBI Volume to Be Infused W Watts Term Description ...
Page 147: ...Local Contacts for Servicing Write your contacts here ...
Page 148: ...11378 2_tm_agilia_sp_eng ...
Page 305: ...Local Contacts for Servicing Write your contacts here ...