29
5.3
Attaching the pump(s)
5.3.1
Attaching to a Pole
5.3.2
Attaching to a Rail
1.
Fold the pole clamp down to the horizontal position:
2.
Unscrew the clamp, attach to the
pole, and screw the clamp until the
pump is fully secured to the pole.
3.
Make sure that the pump is securely
attached.
For more information on installing the pump on a pole, consult the pole’s
Instructions for Use.
Information
When installed on a rolling stand, do not tip over the system more than
5°: it may fall.
Only single pumps can be attached to a bed rail or gurney rail.
1.
Rotate the pole clamp to the vertical position: see section 5.2.2.3,
2.
Unscrew the clamp, attach to the rail,
and screw the clamp until pump is
fully secured to the rail.
3.
Make sure that the pump is securely
attached.