![Fresenius Kabi Agilia SP Technical Manual Download Page 170](http://html.mh-extra.com/html/fresenius-kabi/agilia-sp/agilia-sp_technical-manual_2331626170.webp)
22
4
Fundamentals
4.1
Profiles
4.1.1
Basic Profile
A
profile
defines the device configuration and drug library used for a
group of patients in a given health care environment.
By default, factory settings include only 1 profile (Basic Profile).
Custom profiles can be created and loaded to the pump with Agilia
Vigilant Drug’Lib software. Custom profiles feature a specific pump
configuration and a drug library.
A pump can manage up to 20 profiles:
1 Basic Profile
Up to 19 custom profiles
Information
For pumps used on only one group of patients, we recommend disabling
the ability to select the profile, thus locking the pumps to the selected
profile.
Basic Profile allows programming of an infusion whose settings have not
been pre-defined with Agilia Vigilant Drug’Lib software.
To program an infusion with Basic Profile, choose "Basic Profile" when
selecting a profile.
Basic Profile has the following characteristics:
All infusion settings are to be defined.
The Agilia Vigilant Drug’Lib software's safeguards are
unavailable:
- The infusion is programmed without drug names.
- Limits on drug infusion rates are not included.
Configurations and settings accessible in Basic Profile may not be
suitable for all patient groups and protocols.
Summary of Contents for Agilia SP
Page 139: ...139 17 Spare Parts Catalog Refer to the spare parts catalog of the relevant pump ...
Page 143: ...143 VTBI Volume to Be Infused W Watts Term Description ...
Page 147: ...Local Contacts for Servicing Write your contacts here ...
Page 148: ...11378 2_tm_agilia_sp_eng ...