Chapter 2 Working with DaqLab
55
2.5.7. Temperature
Units
To select between Fahrenheit (
°
F) and Celsius (
°
C) temperature units:
Click
Logger
on the menu bar, point to
Temperature units
and then select the desired units.
2.5.8.
Calibrating the Sensors
Note:
This section reflects software updates made in
DaqLab v1.30.09
and higher. If you are running an
older version of DaqLab, some of the features described here may differ from the features available in
your DaqLab version.
1. Introduction to
DaqPRO
Calibration
The
DaqPRO
ships from the factory fully calibrated. When you first connect your unit to DaqLab, the
software automatically saves these factory calibration settings to two locations:
•
To a file named
factorydefaults.daq
, located in the DaqLab directory on your PC.
•
To
a
.cal
file, located in the
Calibration Settings
folder in the DaqLab directory on your PC. The .cal
file name is your
DaqPRO
’
s
S/N e.g. 123456.cal.
2. Working with the factorydefaults.daq File
DaqLab automatically stores the factory default calibration parameters of the
DaqPRO
the first time the
logger communicates with the software. When you connect additional
DaqPRO
’
s
to the PC for the first,
their calibration parameters are also stored in this file.
As this file must maintain the factory default settings, it is never updated, even after you recalibrate your
DaqPRO
.
As mentioned above, the
DaqPRO
ships fully calibrated. However, replacing the firmware chip or the
internal lithium battery erases these calibration parameters on the
DaqPRO
and resets the logger
’
s S/N
to 0.
When you then connect the
DaqPRO
to DaqLab, the following error message will therefore appear:
Figure 17: Lost calibration error message