Service Manual 1001 5015 / Rev. 4
4:27
4.5.6
Test Weight
Fig. 4:13 Test Weight
Position Settings
This function is for testing of the Wiper, Shutter and Sensor.
Calibrate
– Balance Control; This function can also be reached from the User Menu. Instruc-
tions given on the display.
– Mass Calibration; Instructions given on the display. Use two weights with good ac-
curacy and two decimals. The result is stored in the EEPROM of the TWM.
– Adjust Balance; Should only be used after discussion with Product specialist at
FOSS.
Test Weight Cycle
This function runs a complete Test weight cycle (sequence) but with possibilities to
check part sequences and continue. The raw data will be presented on the display to-
gether with the calculated value. It is possible to catch the sample in the cuvette and
measure on another scale of sufficient accuracy.
AD Values
Continues reading of AD values with an interval of 10 seconds.
Flush
This function empties the hopper content into the drawer.
4.5.7
Sample Transport Module
Position Settings
Possibility to manually position the Cuvette Holder.
Service
Test of sensors and motor. Can be used to run the motor up and down.
Exercise
No use for this option on the field.
Summary of Contents for Infratec 1241
Page 8: ...Infratec 1241 Grain Analyzer Service Manual 1001 5015 Rev 4 ...
Page 34: ...Infratec 1241 Grain Analyzer 2 24 Service Manual 1001 5015 Rev 4 ...
Page 68: ...Infratec 1241 Grain Analyzer 3 34 Service Manual 1001 5015 Rev 4 Fig 3 66 ...
Page 203: ...Infratec 1241 Grain Analyzer Service Manual 1001 5015 Rev 4 5 75 Fig 5 77 Fig 5 78 Fig 5 79 ...
Page 206: ...Infratec 1241 Grain Analyzer 5 78 Service Manual 1001 5015 Rev 4 ...
Page 208: ...Infratec 1241 Grain Analyzer 6 2 Service Manual 1001 5015 Rev 4 ...
Page 209: ...Infratec 1241 Grain Analyzer Service Manual 1001 5015 Rev 4 7 1 7 Schematics N A ...
Page 210: ...Infratec 1241 Grain Analyzer 7 2 Service Manual 1001 5015 Rev 4 ...
Page 212: ...Infratec 1241 Grain Analyzer 8 2 Service Manual 1001 5015 Rev 4 ...