Service Manual 1001 5015 / Rev. 4
4:25
4.5
Manual Tests
To exercise all of the instrument low-level hardware commands. All the tests below
are addressed one by one and performed in manual order.
Fig. 4:11 Manual Tests
4.5.1
Detector
Available options:
•
Set Gain value
•
Read Gain value
•
Read detector temperature
4.5.2
Monochromator
Available options:
•
Set shutter position
•
Read shutter position
•
Set lamp On/Off
•
Set grating motor On/Off
Gain
Pre Gain
Main Gain
Total Gain
1
1
1
1x1
2
1
2
1x2
3
1
4
1x4
4
1
8
1x8
5
1
16
1x16
6
31
1
31x1
7
31
2
31x2
8
31
4
31x4
9
31
8
31x8
10
31
16
31x16
11
1000
1
1000x1
12
1000
2
1000x2
13
1000
4
1000x4
14
1000
8
1000x8
15
1000
16
1000x16
Summary of Contents for Infratec 1241
Page 8: ...Infratec 1241 Grain Analyzer Service Manual 1001 5015 Rev 4 ...
Page 34: ...Infratec 1241 Grain Analyzer 2 24 Service Manual 1001 5015 Rev 4 ...
Page 68: ...Infratec 1241 Grain Analyzer 3 34 Service Manual 1001 5015 Rev 4 Fig 3 66 ...
Page 203: ...Infratec 1241 Grain Analyzer Service Manual 1001 5015 Rev 4 5 75 Fig 5 77 Fig 5 78 Fig 5 79 ...
Page 206: ...Infratec 1241 Grain Analyzer 5 78 Service Manual 1001 5015 Rev 4 ...
Page 208: ...Infratec 1241 Grain Analyzer 6 2 Service Manual 1001 5015 Rev 4 ...
Page 209: ...Infratec 1241 Grain Analyzer Service Manual 1001 5015 Rev 4 7 1 7 Schematics N A ...
Page 210: ...Infratec 1241 Grain Analyzer 7 2 Service Manual 1001 5015 Rev 4 ...
Page 212: ...Infratec 1241 Grain Analyzer 8 2 Service Manual 1001 5015 Rev 4 ...