Service Manual 1001 5015 / Rev. 4
4:41
4.12.3 Status Word 1
Presents the status of the system when the error occurred.
Fig. 4:16
Position in instrument Hardware Error Code in bold:
E.g. "20 - 304/
1ef3
/201/20/0/0/0/450".
Id (hex)
Specification
Note
Source
0001
ADCRunning
ADC communication is OK
ADC
0002
GratingMotorEnable
Grating motor: 0/1 - off/on
DSP
0004
ExerciseModeEnable
Exercise ball bearing in grating
motor: 0/1 - off/on
DSP
00F0
DisplayContrastValue
Contrast of display
PSM
0100
PSM status
0/1 - old/new
PSM
0200
DisplayEnable
Display: 0/1 - off/on
PSM
0400
LampEnable
Monochromator lamp: 0/1 - off/
on
PSM
0800
BacklightEnable
Backlight: 0/1 - off/on
PSM
1000
MUMode
Mode of MU, 0/1 - normal/test
MU
2000
TWMMode
Mode of TWM, 0/1 - normal/test TWM
4000
AbortedCommand
Aborted command indication
MU/TWM
8000
TWMInstalled
Test Weight Module is installed
DSP
Summary of Contents for Infratec 1241
Page 8: ...Infratec 1241 Grain Analyzer Service Manual 1001 5015 Rev 4 ...
Page 34: ...Infratec 1241 Grain Analyzer 2 24 Service Manual 1001 5015 Rev 4 ...
Page 68: ...Infratec 1241 Grain Analyzer 3 34 Service Manual 1001 5015 Rev 4 Fig 3 66 ...
Page 203: ...Infratec 1241 Grain Analyzer Service Manual 1001 5015 Rev 4 5 75 Fig 5 77 Fig 5 78 Fig 5 79 ...
Page 206: ...Infratec 1241 Grain Analyzer 5 78 Service Manual 1001 5015 Rev 4 ...
Page 208: ...Infratec 1241 Grain Analyzer 6 2 Service Manual 1001 5015 Rev 4 ...
Page 209: ...Infratec 1241 Grain Analyzer Service Manual 1001 5015 Rev 4 7 1 7 Schematics N A ...
Page 210: ...Infratec 1241 Grain Analyzer 7 2 Service Manual 1001 5015 Rev 4 ...
Page 212: ...Infratec 1241 Grain Analyzer 8 2 Service Manual 1001 5015 Rev 4 ...