3. Product Description

Signal Quality

Associated status codes

Signal Warning

Temperature Low

Temperature High

Signal Level Low

Signal Level High

Wavelength Low

Wavelength High

Signal Error

Temperature Too Low

Temperature Too High

Signal Level Too Low

Signal Level Too High

Wavelength Too Low

Wavelength Too High

Device Error

Adjustment Missing

No Signal

Device Initializing

3.5.4 Ethernet Interface

In Ethernet mode, the Blackbird OECD transmits the data in the form of UDP
messages.

Communication takes place via IPv6 link-local addresses.

The fos4X multicast address 

ff02::13c

 is used for device recognition.

For communication via the Ethernet interface, fos4X recommends the use of the
fos4X C API. See fos4X C API documentation.

Time synchronization

In Ethernet mode, the Blackbird OECD synchronizes with a PTP master clock with the
following settings:

IEEE1588v2 via UDP via IPv6.

Standard profile E2E.

© fos4X 2020

MDE1261_4-10_1.3.x_5.0_Blackbird

OECD_(12CH, CNO)-en-us

41 / 83

Summary of Contents for Blackbird OECD

Page 1: ...ype Measuring device Product Version 1 3 x Document Version 5 0 Issue Date 2020 02 07 Language en us INSTRUCTIONS FOR USE Blackbird OECD 12CH CNO MDE1261 Product Type Measuring device Product Version 1 3 x Document Version 5 0 Issue Date 2020 02 07 Language en us ...

Page 2: ... Violations will result in the liability for damages All rights reserved in the event of patent utility model or design registration All rights reserved fos4X is a trademark of fos4X GmbH Brand names and protected trademarks are the property of their respective owners The naming of brand names and protected trademarks has only descriptive character Named trademarks are not part of any partnership ...

Page 3: ...ation of Personnel 9 2 4 Intended Use 10 2 5 Residual Risks 10 2 5 1 Explosion Protection 10 2 5 2 Working Environment Wind Turbine 10 3 Product Description 11 3 1 Functional Description 11 3 1 1 Overview 11 3 1 2 Operation 12 3 2 Technical Data 13 3 2 1 Measurement and Sensors 13 3 2 2 Interfaces 14 3 2 3 Optical Data 15 3 2 4 Electrical Data 15 3 2 5 Environmental Conditions 15 3 2 6 Tested Stan...

Page 4: ...4 2 Packaging 50 4 3 Unpacking 50 4 4 Packing 51 4 5 Transport 51 4 6 Storage 52 5 Installation 53 5 1 Important Safety Notes for Installation 53 5 2 Attaching the Measuring Device in the Control Cabinet 53 5 3 Connecting the Interfaces 53 5 4 Connecting the Power Supply 54 5 5 Connecting a Sensor 55 6 Commissioning 57 6 1 Commissioning the Ethernet Interface 57 6 2 Commissioning the CAN Interface...

Page 5: ...leshooting CAN Sub Object 0x2001Sub1 69 9 Maintenance 71 9 1 Preventive Maintenance 71 9 2 Corrective Maintenance 71 9 2 1 Cleaning E 2000 APC Connections 71 9 2 2 Replacing the PWR Connector 72 9 2 3 Replacing the CAN Connector 73 9 2 4 Firmware Update 73 9 3 fos4X Support 76 9 4 Spare Parts and Accessories 77 10 Taking out of Service and Disposal 78 10 1 Taking out of Service 78 10 2 Disassembly...

Page 6: ...afety chapter applies to the entire use of the product See Safety page 9 Safety notes at the beginning of a section apply to the entire section Safety notes before a step apply to the entire following section Safety notes are shown in the operating instructions as follows DANGER Nature and cause of risk Consequences u Measures to avoid the risk The safety notes with the signal words Caution and No...

Page 7: ...of an instruction List Point 1 Point 2 Enumeration with title Identification of Information page 6 Cross reference Bold Text Highlighting of particularly important information Software text Text parameter or input in the software WINDOW Name of a window or button in the software CONTROL ELEMENT Designation of a control element of the product 1 4 Abbreviations CAN Controller Area Network CNO CANope...

Page 8: ... Light Emmiting Diode Max Maximum MTBF Mean Time Between Failures NMT CANopen Network Management OECD Opto Electronic Converter Device PC Personal Computer SMF Single Mode Fiber SNTP Simple Network Time Protocol UDP User Datagram Protocol 8 83 MDE1261_4 10_1 3 x_5 0_Blackbird OECD_ 12CH CNO en us fos4X 2020 ...

Page 9: ...Trained personnel Personnel trained in the operation and safe use of the product Qualified personnel Personnel who based on their specialist training knowledge and experience and knowledge of the relevant regulations are able to assess the work assigned to them and identify possible hazards Assignment of tasks Tasks within the scope of use may only be carried out by the personnel defined below Tab...

Page 10: ... Residual Risks 2 5 1 Explosion Protection The product is not suitable for use in potentially explosive environments 2 5 2 Working Environment Wind Turbine The use of the product in wind turbines requires special risk prevention measures Measures for risk prevention on wind turbines u Obtain and observe safety and work regulations Qualification and training of personnel Personal protective equipme...

Page 11: ...ew TIP Number of channels depends on the variant Figure 1 Overview 1 Mounting plate 2 E 2000 connectors 3 PWR connector 4 Data interfaces The Blackbird OECD consists of the following components Mounting plate Used to attach the Blackbird OECD with 4 M6 bolts or welding studs E 2000 APC connectors 12 sockets for the connection of fiber optical sensors with E 2000 APC connector fos4X 2020 MDE1261_4 ...

Page 12: ...tor CAN MC 1 5 3 GF 3 81 connector for the connection to a CAN network 3 1 2 Operation Up to 12 fiber Bragg grating sensors are connected via fiber optical cables to the E 2000 APC connectors of the Blackbird OECD The Blackbird OECD evaluates the signals of the sensors and provides the wavelengths of the individual sensors via the following interfaces Ethernet CANopen 12 83 MDE1261_4 10_1 3 x_5 0_...

Page 13: ...0 01 pm Warm up time3 20 min 3 dB cut off frequency 500 Hz Internal sample synchrony 300 μs External sample synchrony ETH Precision Time Protocol 300 μs External sample synchrony CAN 4 1 ms 1 Standard deviation at a sampling frequency of 5 Hz 2 At wavelength 1552 nm and signal level 60 3 Time between power up and operation within specification 4 Triggered by CANopen SYNC messages Requirements for ...

Page 14: ...Strain expert Vibration fos4Acc 1D fos4Acc 2D fos4Acc 3D Temperature fos4Temp 3 2 2 Interfaces 3 2 2 1 Ethernet Table 6 Ethernet Interface Property Value Unit Connector RJ45 Transfer rate 10 100 Mbps Max Output rate 1000 Hz Protocol IPv6 PTP protocol IEEE 1588 PTPv2 3 2 2 2 CAN Table 7 CAN Interface Property Value Unit Connector MC 1 5 3 GF 3 81 14 83 MDE1261_4 10_1 3 x_5 0_Blackbird OECD_ 12CH CN...

Page 15: ...annel typical 0 5 1 mW Laser Class DIN EN 60825 1 1 3 2 4 Electrical Data Table 9 Electrical Data Property Value Unit Power supply 10 28 V DC Max Power consumption 15 W Electrical connector MC 1 5 2 GF 3 81 3 2 5 Environmental Conditions Table 10 Environmental Conditions Property Value Unit IP protection class IP20 Maximum operating altitude 3000 m Permitted temperature Transport and storage 40 85...

Page 16: ...ies if the cable connections have the same or a higher protection class 3 2 6 Tested Standards and MTBF Tested Standards Table 11 Tested Standards Test Standard Shock IEC 60068 2 27 10 g in all directions Vibration IEC 60068 2 6 2 g Temperature IEC 60068 2 1 Class 3M6 IEC 60068 2 2 Class 3M6 Electromagnetic Compatibility IEC 61326 1 EN 55032 MTBF Table 12 MTBF Property Value Unit MTBF 100000 h 16 ...

Page 17: ...Unit Length 186 mm Height 141 mm Width 148 mm Total weight 1 6 kg Blackbird OECD without packaging Table 14 Mounting Property Value Unit Mounting 4x M6 3 2 8 Dimensions TIP Number of channels depends on the variant fos4X 2020 MDE1261_4 10_1 3 x_5 0_Blackbird OECD_ 12CH CNO en us 17 83 ...

Page 18: ...3 Product Description Figure 5 Dimensions front view 18 83 MDE1261_4 10_1 3 x_5 0_Blackbird OECD_ 12CH CNO en us fos4X 2020 ...

Page 19: ...3 Product Description Figure 6 Dimensions top view fos4X 2020 MDE1261_4 10_1 3 x_5 0_Blackbird OECD_ 12CH CNO en us 19 83 ...

Page 20: ...he serial number CE marking RoHS marking Indication of the laser class of the product Class 1 Laser Product Power supply 12 24V Power consumption max 15W Optional Device designation KF123 Optional Customer specific ERP number ERP 123456 GTIN P N 4251112345678 3 4 Display and Control Elements 3 4 1 Status LEDs 3 4 1 1 LED Flashing Patterns The Blackbird OECD uses different LED flashing patterns to ...

Page 21: ...ion On LED is continuously on Slow flashing 1 s Fast flashing 1 s Single flash 1 s Double flash 1 s Triple flash 1 s Quadruple flash 1 s 3 4 1 2 LEDs on the PWR connector There are 2 LEDs on the PWR connector fos4X 2020 MDE1261_4 10_1 3 x_5 0_Blackbird OECD_ 12CH CNO en us 21 83 ...

Page 22: ...n the PWR connector STATUS LED Green Yellow Red On System healthy Measurement warning Measurement error Slow flashing Sytem auto adjust System error Fast flashing Identify System not adjusted TIP The STATUS LED flashes quickly every 4 s when in the state On Table 18 STATUS LED and device state Device state Description System healthy Normal operation 22 83 MDE1261_4 10_1 3 x_5 0_Blackbird OECD_ 12C...

Page 23: ... Identify Status is activated by a network command The blink pattern can be used to identify the Blackbird OECD System error Error in the power supply Error in the light source General system error System not adjusted No adjustment for the OE chips loaded Microcontroller defective 3 4 1 3 LEDs on the ETH Connector There are 2 LEDs on the Ethernet connector Figure 9 LED on the Ethernet connector 1 ...

Page 24: ...N Connector 1 Switch for terminating resistor 2 Run LED green 3 Error LED red 4 LED not in use Switch for terminating resistor The CAN connector has a 120 Ohm terminating resistor The terminating resistor can be activated and deactivated via the switch 1 In the right position the terminating resistor is active Run LED Table 20 Run LED on the CAN Connector State Device state Description On Operatio...

Page 25: ... Warning limit reached At least one error counter of the CAN controller has reached or exceeded the warning level Double flash Error control event A guard event NMT slave or NMT master or a heartbeat event heartbeat consumer has occured Triple flash Sync error The sync message was not received within the configured time span for the communication cycle On Bus off Error with the connection to the C...

Page 26: ... interface can be accessed in a web browser within the network via the IP address of the Blackbird OECD Factory setting 198 168 44 44 The web interface is available via the following interfaces Ethernet The web interface has the following functions Monitoring of the system status of the Blackbird OECD and the connected sensors Updating the firmware of the Blackbird OECD Setting the configuration o...

Page 27: ...t the system state CAN Status of the CAN bus Signals Informations about the measurement channels Uptime Operating time of the system System Information Informations about the system Device Type Type designation and version of the Blackbird OECD Serial Number Serial number of the Blackbird OECD Firmware Version Currently installed firmware version fos4X 2020 MDE1261_4 10_1 3 x_5 0_Blackbird OECD_ 1...

Page 28: ...5 2 2 System State The page SYSTEM STATE consists of the following pages CAN STATE SIGNAL STATE HARDWARE STATE SYSTEM REBOOT CAN State Figure 12 CAN State 28 83 MDE1261_4 10_1 3 x_5 0_Blackbird OECD_ 12CH CNO en us fos4X 2020 ...

Page 29: ...N Module State Status of the CAN module CAN Diagnostic Counters Counters of the CAN bus for diagnose purposes Signal State Figure 13 Signal State number of channels depends on the variant Table 25 Signal State Informations about each channel Column Content Channel Number of the channel Wavelength Wavelength of the sensor Raw Signal Signal value of the sensor Reference top and Filter bottom State S...

Page 30: ...te Informations about the CAN interface Column Content Node State Status of the CANopen node CAN Bus State Status of the CAN bus Serial number Serial number of the Blackbird OECD Software State Status of the software Software Software Version Temp Temperature 30 83 MDE1261_4 10_1 3 x_5 0_Blackbird OECD_ 12CH CNO en us fos4X 2020 ...

Page 31: ...re Software Version Temp Temperature LS Number of the light source connected to the channel Table 28 Light Source States Informations about the light source Column Content Number of the light source Laser Power Current power of the laser in percent Laser State Status of the laser of the light source Currents Current of the laser top and of the temperature regulation of the laser bottom Software St...

Page 32: ...tion System Reboot Figure 15 System Reboot A restart of the Blackbird OECD system can be carried out by clicking on the PERFORM SYSTEM REBOOT button 32 83 MDE1261_4 10_1 3 x_5 0_Blackbird OECD_ 12CH CNO en us fos4X 2020 ...

Page 33: ...age VISUALIZATION graphically displays the wavelength curve of the connected sensors The channels can be selected by clicking on the channel at Channel Selection The current wavelengths for all channels are displayed in the Current Values section fos4X 2020 MDE1261_4 10_1 3 x_5 0_Blackbird OECD_ 12CH CNO en us 33 83 ...

Page 34: ...RD FIRMWARE UPDATE System Name Location Figure 17 System Name Location Table 29 System Name Location System name and location Section Content System Name Set name of the system System Location Set location of the system With a click on the button EDIT the editing for System Name and System Location can be activated 34 83 MDE1261_4 10_1 3 x_5 0_Blackbird OECD_ 12CH CNO en us fos4X 2020 ...

Page 35: ...k Current subnet mask Default gateway Current default gateway SNTP Informations about SNTP Enabled Displays if SNTP is on Server IP Current set SNTP server address Syslog Informations about the system log Enabled Displays if the system log is on Server IP Current set server address for the system log With a click on the button EDIT the editing for the settings can be activated fos4X 2020 MDE1261_4...

Page 36: ...played on the CAN MODE page In CAN mode With a click on the button SWITCH TO ETHERNET MODE the Blackbird OECD is rebooted and then works in Ethernet mode In Ethernet mode With a click on the button SWITCH TO CAN MODE the Blackbird OECD is rebooted and then works in CAN mode 36 83 MDE1261_4 10_1 3 x_5 0_Blackbird OECD_ 12CH CNO en us fos4X 2020 ...

Page 37: ...e information about the CAN bus See System State page 28 With a click on the button EDIT the editing for the following settings can be activated Bitrate CANopen Node ID Auto Operational Auto Recover From Bus OFF PDO Transmission Mode fos4X 2020 MDE1261_4 10_1 3 x_5 0_Blackbird OECD_ 12CH CNO en us 37 83 ...

Page 38: ...e page CAN SIGNAL SETTINGS displays the CAN settings for each channel See CANopen Objects page 44 With a click on the button EDIT the editing for the following settings can be activated Offset Scale Factor 38 83 MDE1261_4 10_1 3 x_5 0_Blackbird OECD_ 12CH CNO en us fos4X 2020 ...

Page 39: ... Displays if a password is currently set for the settings Read access without password Displays if read access to the web interface is possible without password With a click on the button CHANGE OR DISABLE PASSWORD the editing for the password settings can be activated fos4X 2020 MDE1261_4 10_1 3 x_5 0_Blackbird OECD_ 12CH CNO en us 39 83 ...

Page 40: ...ality have the following meanings Signal Good Signal quality is good Measurements are possible without restrictions Signal Warning There are one or more warnings See Tab 32 Signal Quality page 40 Measurements are possible The measurements may be inaccurate due to boundary effects and environmental conditions Signal Problem There are one or more errors See Tab 32 Signal Quality page 40 Measurements...

Page 41: ...ernet Interface In Ethernet mode the Blackbird OECD transmits the data in the form of UDP messages Communication takes place via IPv6 link local addresses The fos4X multicast address ff02 13c is used for device recognition For communication via the Ethernet interface fos4X recommends the use of the fos4X C API See fos4X C API documentation Time synchronization In Ethernet mode the Blackbird OECD s...

Page 42: ...tion Part 1 Cabling and connector pin assignment CiA 303 2 Recommendation Part 2 Representation of SI units and prefixes CiA 303 3 Recommendation Part 3 Indicator specification For further informationen see www can cia org Byte order All CANopen messages are in the little endian byte order Node Guarding fos4X recommends Use Node Guarding to monitor the correct operation of the Blackbird OECD For m...

Page 43: ...ms Rebooting the Blackbird OECD A reboot of the Blackbird OECD can be triggered with the NMT command Node Reset 0x81 3 5 5 1 PDOs The Blackbird OECD transmits 4 PDOs PDO1 Table 34 Mapping in PDO1 0 1 2 3 4 5 6 7 2000sub1 Channel 1 2000sub2 Channel 2 2000sub3 Channel 3 2001sub1 Status 1 PDO2 Table 35 Mapping in PDO2 0 1 2 3 4 5 6 7 2000sub4 Channel 4 2000sub5 Channel 5 2000sub6 Channel 6 2001sub2 S...

Page 44: ...x1009 contains from firmware version 2 69 the hardware version number of the Blackbird OECD as visible string The hardware version number has the following format Major Minor Revision 0x100A Manufacturer software version The object Manufacturer software version 0x100A contains the firmware version number of the Blackbird OECD as visible string The firmware version number has the following format M...

Page 45: ...rmed with the offset and scaling factor for each channel See 0x2004 Scale Factor page 48 and 0x2005 Offset page 49 Offset and scaling factor can be configured via the web interface See Web Interface page 26 Error Values Table 39 Error values in 0x2000 Value Description 0x8000 Other sensor error 0x8001 Channel not available on this measurement device 0x8002 No sensor detected on this channel 0x8003...

Page 46: ...re not meaningful in this state 4 Webserver Error The web server is not working The web interface is not available 5 Reserved Bit reserved for future assignment 6 Reserved Bit reserved for future assignment 7 Light Source Error A light source is not working 8 Temperature Error At least one component outside of its permissible operating temperature 9 Signal Low The signal from at least one sensor i...

Page 47: ...it is incremented at each Sync frame 0x2001sub4 Status4 Table 43 Bits in 0x2001sub4 Status4 Bit Name Description 0 11 Reserved Bits reserved for future assignment 12 15 Sync counter Counter bits The counter bit is incremented at each Sync frame 0x2002 Lower Limit of the Measuring Range Description Object 0x2002 contains the lower limit of the measuring range for all channels Each sub object of 0x2...

Page 48: ...t channel Object 0x2003 is read only Sub Objects All sub objects are 32 bit unsigned integers The sub objects contain the 20 bit fix point representation of the wavelength To calculate the wavelength the raw value must be divided by 1048576 Sub Objects for channels that are not present contain the value 0 0x2004 Scale Factor Description Object 0x2004 contains the scale factor for all channels Each...

Page 49: ...e value for the first channel Object 0x2005 is read only Sub Objects All sub objects are 32 bit unsigned integers The sub objects contain the 20 bit fix point representation of the wavelength To calculate the wavelength the raw value must be divided by 1048576 Sub Objects for channels that are not present contain the value 0 fos4X 2020 MDE1261_4 10_1 3 x_5 0_Blackbird OECD_ 12CH CNO en us 49 83 ...

Page 50: ...lackbird OECD is packed in a box with foam inserts One box contains up to 5Blackbird OECD The box has the dimensions 645 x 492 x 216 mm length x width x heigth 4 3 Unpacking Unpacking the the measuring device 1 Open the box 2 Remove the upper foam part from the box 3 Remove the measuring device from the box 4 Check the scope of delivery See Scope of Delivery page 50 5 If parts of the scope of deli...

Page 51: ... device in a different packaging 1 Choose a box so that the measuring device with padding material fits into the box 2 Open the box 3 Pack the measuring device in the box with padding material Make sure that the measuring device is protected against shocks from all sides 4 Close the box P Blackbird OECD ready for transport or shipping 4 5 Transport Transporting the Blackbird OECD u Handle the Blac...

Page 52: ...Store the Blackbird OECD in a dry and clean place u Store the Blackbird OECD in a transport packaging See Packing page 51 u Observe the environmental conditions See Environmental Conditions page 15 52 83 MDE1261_4 10_1 3 x_5 0_Blackbird OECD_ 12CH CNO en us fos4X 2020 ...

Page 53: ... the Control Cabinet Conditions Control cabinet with sufficient space for installation See Dimensions page 17 Control cabinet with permissible environmental conditions See Environmental Conditions page 15 Attaching the measuring device in the control cabinet u Attach the measuring device with 4x M6 screws and nuts TIP Control cabinets from fos4X provide 4 welding studs to attach the measuring devi...

Page 54: ...5 3 GF 3 81 connector to the CAN connector 5 4 Connecting the Power Supply TIP The measuring device has an integrated protection against reverse polarity If the polarity of the power supply is incorrect the measuring device is protected against damage Figure 26 PWR connector Connecting the power supply 1 Switch off the power supply 2 Attach the cable of the power supply to the MC 1 5 2 GF 3 81 con...

Page 55: ...efore connecting TIP The assignment of channels and sensors depends on the application of the Blackbird OECD If available Observe the color coding of the sensors and channels Material and equipment Table 44 Material and equipment Name Type Quantity GTIN Optical connector cleaner Recommended Fujikura OCC A Ultra 1 4251140910209 Connecting a Sensor Figure 27 E 2000 APC connector and socket 1 Locking...

Page 56: ...tive cover 3 on the connector 2 4 Clean the connector 2 with an optical connector cleaner See manual of the optical connector cleaner 5 Close the protective cover 3 6 Connect the connector 2 to the socket 4 P The connector 2 snaps into the socket 4 with a clicking sound P The sensor is connected 7 Repeat the procedure for all sensors 56 83 MDE1261_4 10_1 3 x_5 0_Blackbird OECD_ 12CH CNO en us fos4...

Page 57: ...ce is displayed 2 Click on SETTINGS 3 fos4X recommends Set a password See Set or change the password page 57 4 fos4X recommends Set a system name and location See Settings page 34 5 Set a CANopen Node ID factory setting 4 6 Set the CANopen data rate to the data rate of the CAN bus factory setting 500 7 Set the network settings Web Interface page 26 6 3 Set or change the password Setting or changin...

Page 58: ...Figure 28 No password set Figure 29 Password set 5 If no password is set Click on CREATE PASSWORD 6 If a password is set Click on CHANGE OR DISABLE PASSWORD Figure 30 Setting a password 58 83 MDE1261_4 10_1 3 x_5 0_Blackbird OECD_ 12CH CNO en us fos4X 2020 ...

Page 59: ...New Password 9 Enter the new password in the field at Verify Password 10 Optional Check the box at Read access without password The option Read access without password allows read access to the web interface without password fos4X 2020 MDE1261_4 10_1 3 x_5 0_Blackbird OECD_ 12CH CNO en us 59 83 ...

Page 60: ...page 62 7 2 Opening the Web Interface Opening the web interface 1 Connect the PC to the Ethernet network where the measuring device is located 2 Optional Using a RJ45 cable connect the PC to the ETH connector of the measuring device 3 Open a browser on the PC 4 Enter the IP Adresse of the measuring device in the address bar Factory setting 198 168 44 44 If the option READ ACCESS WITHOUT PASSWORD i...

Page 61: ... on LOGIN The page STARTING WEB APPLICATION is displayed Figure 32 Initialization of the Web Interface 6 Wait until the initialization is complete P The page HOME is displayed Figure 33 Home page fos4X 2020 MDE1261_4 10_1 3 x_5 0_Blackbird OECD_ 12CH CNO en us 61 83 ...

Page 62: ...ptions do not help to fix the problem u Contact fos4X Support See fos4X Support page 76 Table 45 Troubleshooting PWR LEDs Problem Possible cause Solution POWER LED on the PWR connector is not on Blackbird OECD not supplied with power u Check the power cable for damage u Check that the power cable is plugged in u Check that the power supply is switched on 62 83 MDE1261_4 10_1 3 x_5 0_Blackbird OECD...

Page 63: ...se of the sensor 5 Open the page HARDWARE STATE 6 Check the temperature of the components 7 If the temperature of a component is too high Make sure that the temperature regulation in the control cabinet functions properly System not configured u Do a firmware update See Firmware Update page 73 STATUS LED on PWR connector does not light green See LEDs on the PWR connector page 21 System error 1 Che...

Page 64: ...he Blackbird OECD Turn the power supply off and on Signal Warning Wavelength Low or Wavelength High u Make sure that connected sensors are within the measuring range See Measurement and Sensors page 13 u Make sure that only permitted sensors are connected See Measurement and Sensors page 13 Temperature Too Low or Temperature Too High u Make sure to operate within specification See Environmental Co...

Page 65: ...supply off and on Calibration Missing u Contact fos4X Support See fos4X Support page 76 No Signal u Make sure that a sensor is connected u Clean the optical connectors See Cleaning E 2000 APC Connections page 71 u Reboot the Blackbird OECD Turn the power supply off and on Device Initializing u Wait until the initialization is complete 8 3 Troubleshooting Ethernet TIP If the described options do no...

Page 66: ...ontact fos4X Support See fos4X Support page 76 8 4 1 Troubleshooting CAN Connection and LEDs Table 48 Troubleshooting CAN Connection and LEDs Problem Possible cause Solution Cable connection defective u Check the CAN cable for damage Replace damaged cables u Make sure that the CAN cable is connected correctly CAN Interface not configured correctly u Configure the CAN interface See Commissioning th...

Page 67: ...gured correctly See Commissioning the CAN Interface page 57 u Make sure that the Bus Master is configured correctly u Make sure that the bit rate of the CAN bus is configured correctly u Check the CAN cable for damage Replace damaged cables u Make sure that the CAN cable is connected correctly u Do a firmware update See Firmware Update page 73 8 4 2 Troubleshooting CAN Object 0x2000 TIP The number...

Page 68: ...g E 2000 APC Connections page 71 u Reboot the Blackbird OECD Turn the power supply off and on u Check the signal quality See Troubleshooting Signal Quality page 64 Object 0x2000 has value 0x8003 Sensor temporarily unavailable for measurement light source regulates power u Wait until regulation is complete u Reboot the Blackbird OECD Turn the power supply off and on u Check the signal quality See T...

Page 69: ...is u Wait until the automatic troubleshooting is complete u Reboot the Blackbird OECD Turn the power supply off and on Bit 4 of 0x2001sub1 set See 0x2001 Status page 45 Webserver Error u Reboot the Blackbird OECD Turn the power supply off and on Bit 7 of 0x2001sub1 set See 0x2001 Status page 45 Light Source Error u Reboot the Blackbird OECD Turn the power supply off and on Bit 8 or 11 of 0x2001sub...

Page 70: ...00 APC Connections page 71 u Reboot the Blackbird OECD Turn the power supply off and on u Check the signal quality See Troubleshooting Signal Quality page 64 Bit 10 of 0x2001sub1 set See 0x2001 Status page 45 Warning u Check which warnings are pending See Opening the Web Interface page 60 70 83 MDE1261_4 10_1 3 x_5 0_Blackbird OECD_ 12CH CNO en us fos4X 2020 ...

Page 71: ...ance 9 2 1 Cleaning E 2000 APC Connections Figure 34 E 2000 APC connector and socket 1 Locking mechanism 2 Connector 3 Protective cover 4 Socket 5 Dust cap Table 51 Material and equipment Name Type Quantity GTIN Optical connector cleaner Recommended Fujikura OCC A Ultra 1 4251140910209 Cleaning E 2000 APC connections 1 Release the locking mechanism 1 of the connector 2 2 Remove the connector 2 fro...

Page 72: ...00 APC connection is cleaned 10 Repeat the procedure for all connections 9 2 2 Replacing the PWR Connector Table 52 Material and equipment Name Type Quantity GTIN Power supply connector MC 1 5 2 GF 3 81 1 4017918050313 Figure 35 PWR connector Replacing the PWR connector 1 Switch off the power supply 2 Remove the MC 1 5 2 GF 3 81 connector from the PWR connector 3 Remove the power supply cable from...

Page 73: ...N connector 3 Remove the CAN cable from the MC 1 5 3 GF 3 81 connector 4 Dispose of old MC 1 5 3 GF 3 81 connector in accordance with local regulations 5 Attach the CAN cable to the MC 1 5 3 GF 3 81 connector Observe the assignment of the pins High H Low L and Ground G at the CAN connector 6 Connect the MC 1 5 3 GF 3 81 connector to the CAN connector 7 Switch on the power supply 9 2 4 Firmware Upd...

Page 74: ...plete 9 Reboot the Blackbird OECD Turn the power supply off and on P Firmware update complete 9 2 4 2 Firmware Update via fos4X Firmware Flasher Table 54 Material and equipment Name Type Quantity GTIN PC Ethernet connector RJ45 1 Ethernet cable RJ45 1 Software fos4X Firmware Flasher f3 exe 1 Conditions Firmware archive for the Blackbird OECD available IPv6 protocol enabled on the PC Doing a firmwa...

Page 75: ... OECD is not displayed a Make sure that the PC is connected to the ETH port of the Blackbird OECD b Make sure that the correct network adapter is selected c Make sure that the IPv6 protocol is enabled on the PC d Make sure that the PC has a valid IPv6 link local address e Click on in Network Interface 7 Click on in Firmware Archive 8 Select the firmware archive 9 Click on OPEN The firmware archive...

Page 76: ...MWARE b Only if only FLASH BOOTLOADER FIRMWARE is displayed with a green check mark Click on FLASH BOOTLOADER FIRMWARE The status of the update is displayed in a new window 13 Wait until the firmware update is complete 14 Disconnect the PC from the Blackbird OECD 15 If necessary Connect the cable to the ETH connector 16 Reboot the Blackbird OECD Turn the power supply off and on P Firmware update c...

Page 77: ...turer GTIN Power supply connector MC 1 5 2 GF 3 81 Phoenix Contact 4017918050313 CAN connector MC 1 5 3 GF 3 81 Phoenix Contact 4251140902143 Accessories Table 56 Accessories Name Type Manufacturer GTIN Optical connector cleaner OCC A Ultra Fujikura 4251140910209 fos4X 2020 MDE1261_4 10_1 3 x_5 0_Blackbird OECD_ 12CH CNO en us 77 83 ...

Page 78: ... Connector 3 Protective cover 4 Socket 5 Dust cap 2 Disconnect the optical connections a Release the locking mechanism 1 of the connector 2 b Remove the connector 2 from the socket 4 c Attach a dust cap 5 on the socket 4 d Repeat the procedure for all sensors 3 Remove the power cable from the PWR connector 4 Remove the RJ45 cable from the ETH connector 5 Remove the CAN cable from the CAN connector...

Page 79: ...assembled by fos4X or a specialized disposal company 10 3 Disposal NOTICE Risk to the environment due to incorrect disposal of the product Improper disposal of the product may cause damage to the environment u Observe applicable environmental regulations The Blackbird OECD is an electrical appliance and must not be disposed of with household waste Disposing of the Blackbird OECD u Contact fos4X Su...

Page 80: ...ualization number of channels depends on the variant 33 Fig 17 System Name Location 34 Fig 18 Network Settings 35 Fig 19 CAN Mode 36 Fig 20 CAN Bus Settings 37 Fig 21 CAN Signal Settings 38 Fig 22 System Password 39 Fig 23 Firmware Update 40 Fig 24 ETH connector 53 Fig 25 CAN connector 54 Fig 26 PWR connector 54 Fig 27 E 2000 APC connector and socket 55 Fig 28 No password set 58 Fig 29 Password se...

Page 81: ...D on the CAN Connector 25 Tab 22 Factory Settings 25 Tab 23 Home Information on the home page 27 Tab 24 CAN State Informations about the CAN interface 29 Tab 25 Signal State Informations about each channel 29 Tab 26 CAN State Informations about the CAN interface 30 Tab 27 FBG Channel States Informations about each channel 31 Tab 28 Light Source States Informations about the light source 31 Tab 29 ...

Page 82: ...leshooting Ethernet 66 Tab 48 Troubleshooting CAN Connection and LEDs 66 Tab 49 Troubleshooting CAN Object 0x2000 67 Tab 50 Troubleshooting CAN Sub Object 0x2001Sub1 69 Tab 51 Material and equipment 55 Tab 52 Material and equipment 72 Tab 53 Material and equipment 73 Tab 54 Material and equipment 74 Tab 55 Spare Parts 77 Tab 56 Accessories 77 82 83 MDE1261_4 10_1 3 x_5 0_Blackbird OECD_ 12CH CNO e...

Page 83: ......

Reviews: