Calibration
73
2.4.
Date
shows when the water was last calibrated.
2.5.
Confirm
Start?
by choosing
. The calibration procedure will start. The set value of
500 ml will be shown first in the display.
2.6.
Confirm
Exit automatic mode?
by choosing
. This message will only be dis-
played if your automatic feeder is still in automatic mode.
The device now performs two check weighings without the user having to do anything
and shows the result.
2.7.
Then the average value of these two control weighings is shown flashing in the dis-
played line. Confirm with
. You then return to the calibration menu.
2.8.
Date
now shows the current date.
2.9.
Repeat the calibration to check your results.
3. Calibrate other
liquid components
using the same method.
Proceed as follows to calibrate powder components using the calibration scales:
1. Navigate via
> Calibration
to the
Components
sub-menu.
2. In the
MP
menu, you calibrate the milk substitute.
2.1.
Confirm
MP
by choosing
.
2.2.
Set quantity
shows the amount of milk substitute (MP) to be dispensed by the auto-
matic feeder.
2.3.
Runtime
shows the time in which the automatic feeder should dispense the MP.
2.4.
Date
shows when MP was last calibrated.
2.5.
Confirm
Start?
by choosing
.
2.6.
Confirm
Exit automatic mode?
by choosing
. This message will only be dis-
played if your automatic feeder is still in automatic mode.
2.7.
Place the calibration box into the mixer beaker and confirm the prompt
Calibration
box used?
by choosing
.
MP will be dispensed into the calibration beaker.
The device now performs two check weighings without the user having to do anything
and shows the result.
2.8.
Then the average value of these two control weighings is shown flashing in the dis-
played line. Confirm with
. You then return to the calibration menu.
2.9.
Date
now shows the current date.
2.10. Remove the calibration beaker from the mixer beaker.
2.11. Repeat the calibration to check your results.
Proceed as follows to calibrate the cleaning agent using the calibration scales:
1. Navigate via
> Calibration
to the
Components
sub-menu.
2. In the
Flushing agent
menu, you calibrate the cleaning agent.
2.1.
Confirm
Flushing agent
by choosing
.