Forge Medical VasoStat Instructions For Use Download Page 1

VasoStat™ HEMOSTASIS DEVICE 

INSTRUCTIONS FOR USE 

 

CAUTION

:  Rx  only:  U.S.  federal  law  restricts  this  device  to  sale  to  or  on  the  order  of  a  licensed  health 

care practitioner. 

DEVICE DESCRIPTION

: The VasoStat™ Hemostasis Device is used to augment hemostasis through a 

ratcheting pressure mechanism. 

INTENDED  USE

:  The  VasoStat  Hemostasis  Device  is  indicated  for  use  by  medical  professionals  to 

promote  hemostasis  following  catheterization  o

r  other  puncture  into  a  blood  vessel  in  a  patient’s  arm, 

including  radial  artery  catheterization,  arterial  or  venous  line  removal,  hemodialysis,  and  in  patients  on 
anticoagulation therapy. 

CONTRAINDICATIONS

: Allergy to medical adhesives. 

PRECAUTIONS

 

The device is intended for single use only. Do not sterilize and/or reuse this device. 

 

Do not use this product if there is doubt as to whether the product is sterile. 

 

Store in a dark, dry, cool place. Avoid extended exposure to light. 

 

Upon removal of the product from packaging, inspect product to ensure no damage has occurred. 

 

POTENTIAL ADVERSE EVENTS: 

 

Undesired  bleeding  at  the  puncture  site  could  occur  if  care  is  not  used  when  securing  the  device  at 
puncture site 

 

Hematoma 

 

Radial artery thrombosis if the device is tightened excessively without checking for the presence of a 
radial pulse and/or hand perfusion distal to the site of device application 

 

Vascular  thrombosis  if  the  device  is  tightened  excessively  over  a  hemodialysis  access  without 
checking for the presence of a thrill 

 

As in common with skin adhesives, slight irritation of the skin may occur with prolonged use 
 

INSTRUCTIONS FOR USE: 

 

1.  Remove  the  paper  strips  on  the  adhesive 

wings of the 

VasoStat™ Hemostasis Device to 

expose  the  adhesive  surfaces  on  the  base  of 
the device.  Ensure the plunger of the device is 
in  the  up  position.    If  needed,  gently  squeeze 
the  wings  of  the  plunger  together  to move  the 
plunger to the 

‘up’ position. 

 
2.  Align  the  central  plunger  of  the  device  above 

the  vascular  puncture  site  of  the  patient's  skin 
while  the  vascular  sheath  or  needle  is  still  in 
place.    If  needed,  partially  withdraw  the 
vascular sheath or needle  to provide sufficient 
room  for  positioning  and  securement  of  the 
device. 

 

 

3.  Secure the adhesive wings of the device to the 

patient's  skin  with  manual  pressure.    Ensure 
the  adhesive  pads  completely  contact  the 
patient’s  skin.    Remove  the  paper  strips  and 
a

pply  the  Angel  Wings™  adhesive  pads  over 

the  wings  of  the  VasoStat™  device  and 
surrounding skin.

 

 

 
4.  While  stabilizing  the  base  of  the  device  with 

index  and  middle  fingers,  depress  the  plunger 
of  the 

VasoStat™  device  with  gentle  thumb 

pressure  until  it  is  just  contacting  the  patient's 
skin.  

 

Reviews: