DUALEX SCIENTIFIC
+
TM
FORCE-A
User’s manual v.3.1
14
II.6.c.
Measurement
Measurements can be performed by alternative ways:
-
Triggering
:
Automatic (by default): at every closure of the clip or a press on the button
, a measurement is
saved.
Manual: the recording of the measurement is only made once the button
is pressed.
Tuning: this function re-adjusts the parameters to allow the automatic mode, depending on the clip
aperture. Use it if automatic mode doesn’t work.
-
1 Side/2 Sides
:
1 Side: at every closure of the clip or a press on the button
, a measurement is recorded
(following the triggering previously selected). At every measurement, values Chl, Flav, Anth and
NBI
of the side
(adaxial
or
abaxial side) are displayed.
Mean 2 Sides: 2 measurements are needed to get values for one leaf (adaxial and abaxial side).
Chl = mean of 2 values
Flav = sum of 2 values
Anth = mean of 2 values
NBI = Chl / Flav.