D40 Owner’s Manual
EN-12
Testing Your Blood Glucose
Test Strip Appearance
Absorbent Hole
Apply a drop of blood here.
Confirmation Window
An area where you confirm
if enough blood has been
applied to the absorbent hole
in the strip.
Test Strip Handle
Hold this part to insert
the test strip into the slot.
Contact Bars
Insert this end of the test strip
into the meter. Push it in firmly
until it will go no further.
Inserting a Test Strip
Insert the test strip as shown in the illustration.
Important!
The front side of test strip with FORA Logo should face up when
inserting test strip. Test results might be wrong if the contact bar is not
fully inserted into the test slot.
Preparing the Lancing Device
1. Remove the cap.
2. Insert a new lancet firmly into the white lancet holder cup.
3. Remove the protective disk on the lancet.
Hold the lancet firmly in place and twist off the protective disk.