background image

Operation and Service Manual

Follett Advantage™  

FZRADV25-LB Freezers 

01401397R00

H E A LT H C A R E

TM

801 Church Lane • Easton, PA 18040, USA

Toll free (877) 612-5086 • +1 (610) 252-7301

www.follettice.com/healthcare

Following installation, please forward this manual 

to the appropriate operations person.

Summary of Contents for Advantage FZRADV25-LB

Page 1: ...FZRADV25 LB Freezers 01401397R00 H E A LT H C A R E TM 801 Church Lane Easton PA 18040 USA Toll free 877 612 5086 1 610 252 7301 www follettice com healthcare Following installation please forward th...

Page 2: ...mote Alarm Connection 6 Controller Operation 7 Locking Unlocking of the Keypad 8 Viewing Adjusting the Set Point 8 Powering the Equipment ON 8 Controller Set Up 8 Defrosting 9 Door Switch 9 Cleaning 1...

Page 3: ...than 30 off vertical during uncrating or installation After uncrating and removing all packing material inspect the equipment for concealed shipping damage If damage is found notify the shipper immed...

Page 4: ...ensure proper ventilation of the refrigeration system The back of the freezer should have 2 inches of clearance to allow for power cord clearance and ventilation on unit Electrical Specifications 115...

Page 5: ...nditions other than those specified herewith The equipment is secured to a wooden pallet base wrapped in a plastic film and packaged into a three waves carton box The equipment must be handled using a...

Page 6: ...N m Fig 2 2 4 Attach wall anchor brackets to wall using appropriate hardware and installation method suitable to application and local codes Fig 2 1 2 Remote Alarm Connection These units are equipped...

Page 7: ...improper system operation and or severe mechanical damage Changing operating parameters may result in Follett Products LLC voiding equipment warranty Freezers are pre programmed with a 25 C 13 F set p...

Page 8: ...oller unlocked press and release the SET key The current Set Point will populate Use the UP DOWN arrows to the desired Set Point and again press the SET key to store Powering the Equipment ON Locate t...

Page 9: ...efrost cycle is 15 minutes If this occurs a dFd alarm will appear Upon termination of the defrost cycle the compressor condenser fan and evaporator fan will remain off for 60 seconds to allow for the...

Page 10: ...10 Follett Advantage FZRADV25 LB Freezers Alarms...

Page 11: ...he glass door and exterior parts may be cleaned with a soft cloth window cleaner or other non abrasive cleaner Annual Cleaning Removal of dust and other particulates from air intake areas and the cond...

Page 12: ...id Low Psi Liquid High Psi Vapor Low Psi Vapor EVAPORATOR FAN EVAPORATOR DEFROST UNIT CONDENSING UNIT EVAPORATOR TXV HOT GAS SOLENOID HOT GAS LINE Component Details Model Gas Type Gas Charge Low side...

Page 13: ...f power to unit and unplug 2 Remove fa ade hold down screws rotate fa ade and set safety latch 3 Disconnect terminal from back up battery 4 Carefully cut wire ties to provide slack in control wires 5...

Page 14: ...ller 1 Firstly make sure that the main switch is turned off and the controller is not powered The controller has removable screw connection terminals blocks so unscrew at the maximum openings the scre...

Page 15: ...ng key is correctly inserted into the controller terminal blocks Copy the parameters from the controller to the programming key 1 After the above steps 1 2 Press and hold the Copy key button until the...

Page 16: ...and press firmly in place 4 Open and close door checking for proper gasket seal without pinching against freezer Modular Refrigeration System Follett Products LLC Modular Refrigeration System allows...

Page 17: ...3 7 4 8 Black White Black White Black White Black White Black White Black White Battery V N L Power supply 100 240Vac V Low Battery AC OK Output 12Vdc Battery 12V Brown Blue Ground Red Black Red Black...

Page 18: ...18 Follett Advantage FZRADV25 LB Freezers Replacement Parts...

Page 19: ...6 1 21 Main Switch Kit 01394238 1 22 Power Cord 01316140 1 23 Top Protection Grid 01398155 1 24 Probes Kit 01394220 1 25 Safety Fan 01394345 1 26 Battery Backup Protection Relay 01394394 1 27 Controll...

Page 20: ...cts LLC will not be liable for any consequential damages expenses connecting or disconnecting charges or any losses resulting from a defect of the machine For full warranty details visit our website w...

Reviews: