leakage sensor) and go back to a ready-to-use status. If the leakage occurred during 
a test cycle, after pressing “Leakage resolved” Mastertronic will immediately be 
followed by an automatic empty procedure, and then also a rinsing procedure. 

Maintenance 

Change external hoses 

We recommend to change the external hoses every 8 months, 

and it very important 

to use exact the same Inner Diameter as original (3.2mm) and same length as 
we provide from start, 2 m for all the external hoses. 

NEVER change the length 

nor the diameter of the external hoses! 

Change hoses in pump 

We recommend to change the hose in the stepper  every 8 months,  and  the hoses 
in pump A,B,C every 12 months. Put the needle in ”maintenance position” , when 
changing the stepper hose, and in ”standby position” when changing hoses in pump 
A-C. For tips on how to change the hoses check out www.Focustronic.net for a video.

 

Recalibration 

Calibration is to be done around 8 weeks. The machine will give you an alert when  
its time to calibrate, as a reminder.  

Descale/Clean the reaction chamber 

Due to the reagents containing colour dyes and also carbonate deposits, its 
advisable to make regular cleaning and descaling of the reaction chamber. The 
interval depends on the measure frequency, but around every 4 weeks is a good rule 
to keep consistent results every time.   

Descaling 

Make sure the machine is in standby mode and the reagent chamber and 
hoses are empty.  

Normally 

 its totally emptied as all test cycles end with 

an empty phase and also after a restart Mastertronic starts with an empty 
program. 

But

 If there is, for some unpredictable reason, despite this water 

in hoses and/or chambers, press ”

Empty Chambers and Hoses

” from the 

app menu.  

Open the app and press ”

Descaling Program

” 

Summary of Contents for Mastertonic

Page 1: ...1 User manual English v 3 2...

Page 2: ...er pump 12 Pump A RO water pump 12 Syringe 13 OLI calibrate 13 Test pump syringe accuracy 13 Stepper pump Sample water pump 13 Pump A RO water pump 13 Syringe 14 Dashboard 14 Parameters menu 15 Mainte...

Page 3: ...ents Focustronic has no responsibility of any kind of injury or damage related to above mentioned items Use Maintenance position of the needle when doing certain maintenance Change hoses in pump 21 Re...

Page 4: ...first time there is a machine Mastertronic that takes care of these crucial parameters for you and therefore raises the level of reefing to an outstanding level We also added the Calcium and Magnesium...

Page 5: ...n Mastertronic is a service to make Mastertronic 100 complete and superior but should never be instead of a pH electrode based KH test Alkatronic In MT the alkalinity test is limited to 1 scheduled te...

Page 6: ...r than the original 3 2mm so when you replace these hoses use the same length and diameter as the original Sample Water IN Use the included hose labeled 1 and on one end attach the microfilter see pic...

Page 7: ...power on the unit yet If this is your fi rst Focustronic product and have no account yet start with step I III users with existing accounts can start from point 1 I Download the Focustronic All in On...

Page 8: ...router and therefore also now online 13 Watch the LCD screen If it does not display the latest software it will now automatically download that As long as the machine is online all the new updates wil...

Page 9: ...nt vial to the carousel it will automatically move to the correct position see below Its also possible to control the carousel from the app see picture NEVER move the carousel with your hands That wil...

Page 10: ...ose Vial position is labeled by a number see pic and the active position is slightly to the left If you for instance assign a reagent to vial 3 the carousel will move so the vial 3 position in the car...

Page 11: ...e number of tests before needing to refill Note For Nitrite you don t have to assign any other reagents Mastertronic uses same reagents for nitrite as you have assigned for nitrate Example If you for...

Page 12: ...p actions will start and do NOT press any other buttons until all pumps have stopped 5 Water has now been delivered in the measuring cylinder 6 Read water amount in the cylinder Enter in the app with...

Page 13: ...uration of the water which is a very good indicator of the overall organic load and decomposition levels With this parameter you have a very good idea of when to replace any carbon media or make other...

Page 14: ...est Pump Accuracy again and follow instructions again from step 3 7 When you are done finish the session by pressing Empty Hose The app will not allow you to leave the page before you press this of se...

Page 15: ...set up hereby entering the heart of the app the parameters menu There is one parameter menu for each parameter so all settings here are only valid for the chosen parameter Graph Latest measurement Tab...

Page 16: ...astertronic to keep correcting corrections Assign Solartronic Coming soon Here you assign the Solartronic to obey from the results of the parameter measured by Mastertronic Typically its the nutrients...

Page 17: ...logics above will make any changes to the assigned Dosetronic channel it will alter the whole channel in terms of 10 change is default but can be customised here When coming to nutrient additions lik...

Page 18: ...s the nitrite is of very low concentration 0 02 mg l so this is negligible and thus does not have this effect on the final nitrate value But in some special situations like an immature tank with slow...

Page 19: ...s that will in a case of an unexpected overfill situation lock down the machine See also this section Rinsing of Reaction Chamber In all normal usage situations rinsing and emptying is of course compl...

Page 20: ...ally never used as all algorithms and failsafes always make sure the syringe is empty So this function is if something unpredictable happens and will take the syringe to the bottom position Empty Wast...

Page 21: ...when changing hoses in pump A C For tips on how to change the hoses check out www Focustronic net for a video Recalibration Calibration is to be done around 8 weeks The machine will give you an alert...

Page 22: ...table reason some water in the hoses and or chambers press Empty Chambers and Hoses from the app menu Use a Q tip cotton bud and soak that with ethanol its ok to use for instance non gel based hand al...

Page 23: ...ly and you shall receive a notification You will have to dry the machine and then press Leakage resolved and Mastertronic will resume For drying properly do this 1 Dry up the visible water that is on...

Page 24: ...sly wrong Like always very very low values then sensor may need to be dried some more and then proceed with the additional steps below A Power off unit B Lift up the chambers leaving a small gap for f...

Page 25: ...factor Now do a test with the manual test kit also on your tank and correct the measured value with the factor Now this value is very likely to be the true value This value should be close to a fresh...

Page 26: ...ess than 8 ml in the reagent vials Then needle will not get any fluid Have you had a leakage last day s Then light sensor might have been humid and need drying Do a light sensor test maintenance secti...

Page 27: ...e activated carbon etc You calibrate the OLI against some RO Distilled water see calibrate section We recommend to recalibrate OLI every 4 weeks Summary of settings in assigned DT ST PT Update Dosetro...

Page 28: ...ocated in device list B then use Dashboard B If you haven t moved MT to your tank section B in the pic then MT is as default located in dashboard A and device list A By press Device list and choose yo...

Page 29: ...If not its either MT is not calibrated correct or issue with the reagent see point 1 3 Concerning ICP and PO4 its also worth to take note ICP measures all phosphorus not just the inorganic PO4 That me...

Page 30: ...measure manual more than maybe once a month now with MT you measure that very often and just see variations you didn t know before But if you have proved Mg actually is too high and MT is correct cali...

Page 31: ...Connectivity Cloud App Update procedure Wifi Automatic update through cloud Software iOS app Android app wifi based Standalone Yes Dosetronic compatible Yes Wireless communication with Dosetronic mak...

Page 32: ...mg l Ca2 6 mg l KH 0 10 dKH I 0 005 mg l Resolution NO3 0 01 mg l NO2 0 001 mg l PO4 0 001 mg l Mg2 1 mg l Ca2 1 mg l KH 0 01dKH I 0 001 mg l Level of detection NO3 0 05 mg l NO2 0 001 mg l PO4 0 003...

Page 33: ...10 11 1 Pump A 2 Pump B 3 Pump C 4 Stepper 5 Needle stabilisator 6 Syringe 12 Author Jonas Roman Cred to Pete Bird for language corrections 7 Reaction chamber 8 Rinsing chamber 9 Reagent carousel 10...

Reviews: