8.0.7 PD

Protocol trigger

Switch the PD communication switch to ON, Enter the PD protocol trigger mode.
After exiting the PD trigger, switch the PD communication switch to OFF.

Take the picture as an example. The picture shows a message sent by a

charger.There are 6 gears in total, of which gears 1, 2, 3, 4, and 5 are fixed
voltage gears.The sixth gear is the adjustable voltage gear (PPS).

When the left dot stays on Gear,The gear can be switched by the left and

right keys. When the gear is switched to PPS gear,Then you can switch the

step voltage by clicking the middle button,After selecting the step voltage,
Through the left and right keys (left minus right plus),Decrease/increase
voltage.

8.0.8 PD

Protocol conversion

This function is used for only QC2.0 charger,But want to supply power to PD

appliances.Before use,Switch the PD communication switch to ON,Then enter the PD
protocol conversion mode, after entering,Plug in PD appliances,You can perform
PD fast charge.

In this mode,Click the middle button and use the left and right buttons to

change the maximum power of packets sent by the PD.When changing power,

Be

careful not to exceed the charger power to avoid unnecessary damage

.

After

changing the power, you must click the middle button to confirm.

Set 5V when no device is connected,Avoid high-voltage damage to mobile

phones that do not support high-voltage when plugged in.

QC2.0 only B type charger supports 20V trigger,So when the PD appliance

requests 20V voltage,The tester will detect whether the charger successfully
triggers QC2.0-20V,If it does not reach 20V,The tester will cancel the 20V gear,

And resend the Caps broadcast.

In addition,Some PD appliances will change the D+ and D- voltages

whencharging,Cause QC2.0 to trigger an exception,This type of PD appliances cannot

usethis function to charge.

Summary of Contents for FNB48

Page 1: ...1 FNB48 User Manual V0 6...

Page 2: ...charge trigger 5 3 0 5 Wire identification class 6 3 0 5 Miscellaneous 6 4 0 Structure appearance 7 5 0 Technical index 10 6 0 Main page 11 6 0 1 Concise page 11 6 0 2 Record page 12 6 0 7 Fast charge...

Page 3: ...8 0 7 PD Protocol trigger 21 8 0 8 PD Protocol conversion 21 8 0 9 VOOC WARP Constant voltage trigger 22 8 0 9 SVOOC Trigger 22 9 0 Charging tool 23 9 0 1 PD Listener 23 9 0 2 Read E Marker cable 25 9...

Page 4: ...l situations Please read the relevant content carefully and strictly abide by all warnings and precautions 2 0 Safety Precautions 1 Do not connect the monitoring interface to a power supply exceeding...

Page 5: ...ch 3 Touch switch 3 0 3 Voltage and Current 1The highest six digit display of voltage current and power the highest resolution is 0 00001 V A W 2 Record the minimum maximum and average values of volta...

Page 6: ...e trigger and automatically close the trigger when the time comes 3 0 5 Wire identification class 1 The internal resistance measurement of the wire by the differential pressure method 2 E Marker Cable...

Page 7: ...3 input monitoring port Micro USB 5 PIN female socket 4 output monitoring port TYPE C 24 PIN female socket 5 PD communication switch 6 output monitoring port USB A 9 PIN female 7 touch switch BACK but...

Page 8: ...wer 0 156W 0 00001W 0 5 2 Load equivalent internal resistance 0 9999 9 0 0001 0 5 2 D D voltage 0 3 3V 0 001V 1 0 2 Equipment temperature 1 1 2 3 1 1 2 4 Capacity 0 9999 99Ah 0 00001Ah Energy used 0 9...

Page 9: ...alid 6 0 1 Concise page Description Only the three key parameters of voltage current and power are displayed indicates the direction of current This page can change the display direction Instructions...

Page 10: ...re are two text boxes in the lower right corner the first 1 0 from top to bottom is the setting menu recording offline recording time in hours it is highlighted when recording otherwise it is grayed o...

Page 11: ...cols stop triggering the charger will restart The dark blue below the trigger timing is the load resistance The first column of the bottom white table is D D voltage The second column is the current c...

Page 12: ...e 3 Record offline curve Mode 4 High speed voltage ripple AC coupling Instructions 1 Long press the left button time base minus 2 Click the middle button start pause drawing the curve Except in mode 3...

Page 13: ...age and current value as the reference value Measurement procedure 1 Connection mode charger FNB48 constant current load the current is adjusted to about 0 5 1A and record the reference value 2 Connec...

Page 14: ...ameters from left to right are group number capacity energy the selected group is displayed in green and the lower left corner is the selected group Statistics time the lower right corner is the group...

Page 15: ...eft right button Each item is explained below 1 Group is the statistical group selected for calculation This instrument can be selected from 1 10 groups statistical time capacity energy It is displaye...

Page 16: ...CP SCP SamsungAFC trigger VOOC DASH constant pressure mode PD2 0 3 0 trigger and QC2 0 PD2 0 protocol conversion Note Once you enter the quick charge trigger detection interface All operations must be...

Page 17: ...d continue testing During the test it is forbidden to connect to any equipment at the back end It does not respond to any key operation during the test If you want to exit during the test please unplu...

Page 18: ...Click BACK to return Press the left right keys to quickly decrease increase the voltage 8 0 4 Huawei FCP trigger The operation method is the same as QC2 0 trigger 8 0 5 Huawei SCP trigger The operati...

Page 19: ...re use Switch the PD communication switch to ON Then enter the PD protocol conversion mode after entering Plug in PD appliances You can perform PD fast charge In this mode Click the middle button and...

Page 20: ...method is the same as QC3 0 trigger 8 0 9 SVOOC Trigger SuperVOOC requires a load greater than 500mA on the back end to deceive And SuperVOOC only has a voltage of 10 5V Therefore you can only press B...

Page 21: ...ovide external power Use the PD listener function Need 2 C C cables Connect the charger and PD electrical appliances from the Type CIN interface and Type COUT interface respectively When the connectio...

Page 22: ...for the gear required for charging for example it is 6 gears in this example request one of them indicates the direction of data transmission It means that this message was sent by the PD consumer to...

Page 23: ...the Type C interface cannot be used for power supply PC port USB A Micro USB interface can all be used for power supply The PD communication switch needs to be turned ON After entering this function F...

Page 24: ...uch thread in hand Only FNB48 tester and a C C cable But want to trigger VOOC WARP flash charge At this time the function of simulating DASH cable can be started And use the C C cable to connect to th...

Page 25: ...the standby screen brightness Adjustable range 0 20 level when set to 0 Enter the standby state and turn off the screen directly 10 1 3 Standby time Set the standby time the last time you operate the...

Page 26: ...settings except the recorded data 10 2 Settings menu Record Set the configuration required for logging data 10 2 1 Curve recording time Set the recording time of the voltage and current curve Maximum...

Page 27: ...Off by default 10 4 Settings menu system You can view information such as the operating times of the instrument SN serial number and software version 11 0 Upgrade firmware instructions 1 Open the FNIR...

Reviews: