Operating instructions digital oval gear meter
9. Troubleshooting
Problem
Cause
Solution
LCD display: no display
Batteries do not contact properly Check batteries for good contact
Accuracy of measurements is
insufficient
K FACTOR wrong
Meter operates below admissible
min. flow rate
Check valid K FACTOR.
See section 6.3
Increase flow rate to acceptable
level
Reduced flow rate, no flow at all
Oval gear blocked
Clean measuring chamber
Display flashing „Err 1“
Data stored in electronic memory
are damaged
Sorry, irreparable
Short displays of „Err 2“
Limited error reading data
(usually when exchanging
batteries)
Control circuit board
automatically switches on and off
to restore proper operation