Evaluating results
Before a result is evaluated, make sure that the measurement is
valid and no error message is present (see page 2-2).
Result details
The following information can be displayed:
- Measurement Task, Measurement Channel, Measurement Range,
- Sensor Type, Date/Time, Status
- Statistical Data (for overall values only)
• Press MENU and click on 'Result Details'.
Characteristic overall value - single values
Up to two overall values are simultaneously displayed in the
result screen. The RPM appears in the lower area. If the threshold
values are exceeded during the measurement, these appear in the
field below the respective measurement value (see below). Also,
the respective LED lights up next to the display (see pages 2-2).
Thresholds can be displayed as an absolute value or as a difference
to the measurement value.
For vibration parameters with more than two overall values*, press
the navigation key up / down to display the other values (see
'Display Setup', page 4-17).
* e.g.: Vibration parameters with 6 overall values:
1. rms value
2. 0-p value
3. p-p value
4. 0-p calculated (= RMS x
√
2)
5. p-p calculated (= 0-p x 2)
6. Crest factor (s. P. 5-1)