2-20
Getting started - Basic settings
Transducer parameters
for VIB 6.122 accelerometer
*'Show' appears for factory-defined
transducers whose parameters cannot
be changed.
1
only for VIBREX / VIBRONET
² only for 'Voltage' signal type
Transducer parameters
To check the parameters for transducers that have already been set
up, proceed as follows:
• Select the transducer.
• Press the MENU key.
• Click on 'Show'* or 'Edit' to open the parameter screen (see
below).
Meas. quantity
Measurement quantity of the sensor: vibration acceleration,
velocity, displacement, current, RPM, jump, user-defined quantity
Signal type
Signal type of the sensor (e.g. LineDrive, ICP, voltage)
Input voltage range
±3V / ±8V
1
/ ±30V²; only applies for transducers of 'Voltage',
Current', 'ext. 5V converter' and 'VIBREX/VIBRONET' signal types.
Quantity (user def.)
Identification for a user-defined measurement variable.
Unit (user. def.)
Unit for the user-defined variable.
Precision
Accuracy in decimal places.
Coefficient a4, a3, a2
Parameters for linearizing a sensor's non-linear characteristic line
(e.g. MNS12 proximity probe - VIB 6.640).
Sensitivity
The sensitivity of the sensor determines the signal level.
Offset
Offset of the sensor
Linear from / to
The linearity range of the sensor and the filter setting of the
Summary of Contents for VIB 5.310