SMFT-1000
Users Manual
12
Terminals/Test Leads
Test leads are connected and stay in place (Keep the Leads) throughout the test.
Table 6
shows the input terminals.
XW
Warning
To prevent possible electrical shock, fire, or personal injury, do not use test leads
in CAT III or CAT IV environments without the protective cap installed. The
protective cap decreases the exposed probe metal to <4 mm. This decreases the
possibility of arc flash from short circuits.
The IR (infrared) port allows you to connect the Tester to a computer and download the test
data with
TruTest™ Data Management Software
documentation. With this software, you can
collect, organize, and display the test data. See
for additional
information on using the IR port.
Table 6. Terminals
Item
Description
Red Socket (PV [+] V input)
Blue Socket (PV [-] COM input)
Green Socket (Earth)
Yellow Socket (R
PE
, Clamp input common)
IR Data Port
2
1
5
1
2
3
4