RUSKA 7252
Users Manual
6-12
glb46.bmp
Figure 6-5. Zeroing in Progress
6.
Press PREVIOUS to return to the Main Menu.
Absolute Instruments
A high capacity vacuum pump must be connected to the rear reference port connector.
Units that were purchased with the Vacuum Sensor option, have an internal vacuum
sensor mounted internal to the primary sensor housing that monitors the vacuum level
when the instrument is being zeroed in the absolute mode. This sensor reading will
display a value of 350 mtorr when the vacuum level is at or greater than 350 mtorr.
1.
Select the desired pressure units using the
Units
key. You can change both the unit
of measure that the Calibrator is displaying Pressure and the unit of measure for the
vacuum sensor. It is not required to have the Calibrator displaying pressure in the
same unit of measure as the vacuum sensor. To change the unit of measure for the
Vacuum sensor, press the
Unit
key and then the
Vacuum
[
F2
] key.
2.
Enter the Calibration
screen by selecting
[MODE] MENU | CALIBRATE
.
3.
Select
Zero
[
F1
]. Do
not
press the Calibrate
button. When initiating the zeroing
command in the Calibrator, it will cycle the internal valves allowing the external
vacuum pump to pull a hard vacuum on the RUSKA sensor. It is recommended that
the sensor be pulled down to a vacuum of less than 200 mTorr (preferably less than
100 mTorr) in order to obtain a proper zero. The Calibrator can be zeroed at higher
vacuum levels, however, the uncertainty of vacuum sensors tend to increase at higher
vacuum levels, therefore, the uncertainty in the Calibrator would also increase if
zeroed at higher vacuum levels.
4.
Wait for the vacuum level to stabilize as read by the internal vacuum sensor.
5.
Verify that the primary sensor pressure and temperature are stable.
6.
Press
OK
[
F6
] when stable.
7.
Do not disturb the instrument while zeroing is in process.
8.
Wait for the zeroing procedure to finish.
9.
Press
PREVIOUS
to return to the Main Menu.
Summary of Contents for RUSKA 7252
Page 3: ...7252 Change Language Hold key for 5 seconds enter mode ...
Page 4: ......
Page 10: ...RUSKA 7252 Users Manual vi ...
Page 12: ...RUSKA 7252 Users Manual viii ...
Page 20: ...RUSKA 7252 Users Manual 1 6 ...
Page 40: ...RUSKA 7252 Users Manual 3 6 ...
Page 72: ...RUSKA 7252 Users Manual 4 32 ...
Page 114: ...RUSKA 7252 Users Manual 6 22 ...
Page 118: ...RUSKA 7252 Users Manual 7 4 ...