Maintenance
Calibration
6
6-11
Zeroing
The zeroing procedure is performed to correct for system zero shift and does not require a
full calibration. The most important requirement for performing a valid zeroing procedure
is to guarantee that there is not a pressure differential between the sensor’s test port and
case reference.
If during the zeroing procedure, the message
“Mechanical Zeroing Needed”
appears,
the sensor photocell may need to be zeroed. See Chapter 6, Sensor Photocell Zeroing, for
more information.
The zeroing screen presents several pieces of information.
The screen will display the current status of the pressure sensor Stability and
Temperature.
If any of the above are unstable, then the system will delay until stability is achieved.
Pressing
OK
[
F6
] will bypass this wait period.
Note
Bypassing this wait period can have a negative effect on the zeroing
procedure.
Gauge and Vacuum (Negative Gauge) Instruments
1.
Verify that the Reference Port is open to atmosphere.
2.
Enter the Calibration screen by selecting [
MENU] MENU |
CALIBRATION
.
3.
ZeroAll
— Selecting,
ZeroAll
will perform a zeroing operation for both sensor A
and sensor B.
Zero
— Selecting
Zero
will perform a zeroing operation on the selected Sensor. Press
Sensor to change the selected sensor.
Do NOT press the
Calibration
button.
glb45.bmp
Figure 6-4. Zeroing Menu
4.
Do not disturb the instrument while zeroing is in process.
5.
Wait for the zeroing procedure to finish.
Summary of Contents for RUSKA 7252
Page 3: ...7252 Change Language Hold key for 5 seconds enter mode ...
Page 4: ......
Page 10: ...RUSKA 7252 Users Manual vi ...
Page 12: ...RUSKA 7252 Users Manual viii ...
Page 20: ...RUSKA 7252 Users Manual 1 6 ...
Page 40: ...RUSKA 7252 Users Manual 3 6 ...
Page 72: ...RUSKA 7252 Users Manual 4 32 ...
Page 114: ...RUSKA 7252 Users Manual 6 22 ...
Page 118: ...RUSKA 7252 Users Manual 7 4 ...