PG9602™ OPERATION AND MAINTENANCE MANUAL
© 2011 Fluke Calibration
Page
84
<6READY>
to view the values of
Ready/Not Ready
status criteria (see Section 3.4). These
include:
<1Pposition>:
Piston position
Ready/Not Ready
limits (see Section 3.4.1). The value is
displayed as a band in millimeters about mid-float position (
±
mm). (see Section
<2vac>:
Maximum reference vacuum pressure limit when operating in absolute by
vacuum mode. The value is displayed in Pascal (Pa).
<7vac>
: to view the reference vacuum variable source choice and the current value. The
value is displayed in Pascal (Pa).
3.11.3 <3edit>
PURPOSE
To edit an existing SETUP file and/or to create a new SETUP file.
OPERATION
Note
•
SETUP file #1 is the factory default file and it cannot be edited.
•
See Table 16 for a listing of the source choices for each variable.
To edit an existing SETUP file or create a new SETUP file, press
[SETUP]
and
select
<3edit>
. The display is:
1.
Entry field for number of SETUP file to be edited or created.
Defaults to SETUP file currently selected. Must be a number
from 2 to 10.
Edit SETUP file:
#02
Enter the desired SETUP file number and press
[ENT]
.
If a valid SETUP file number is entered, the display is:
Edit #02 latmP 2ambT
3%RH 4PCT 5g 6READY
7vac
Selecting an item leads to the variable source choices menu for that item. The cursor is on the
source choice that is currently selected in that SETUP file number. Select the desired
variable source. The selection causes the variable choice selection to be made and returns
to the edit menu. From the variable source choice menus, pressing
[ENT]
or
[ESC]
returns
to the edit selection menu. This allows easy, discreet movement between variable source
choices when editing a SETUP file.
Select
<1atmP>
to specify the atmospheric pressure variable source for the pressure values
that are used to calculate ambient air density and to add to gauge pressure measurements to
calculate absolute pressure in absolute by addition of atmosphere mode. Selecting
<1atmP>
offers the atmospheric pressure variable source choices:
<1internal>
Use real time readings from PG9000’s on-board barometer.
<2normal>
Use a fixed value of 101.325 kPa (14.6959) psi.
<3user>
Use a fixed user defined value. If
<3user>
is selected, the user value must be
entered, in the current pressure unit of measure (see Section 3.10.3).
<4COM2>
Use real time readings from the device connected by RS232 to PG9000’s COM2
port. The external barometer must be properly set up to communicate with PG9000
(see Section 3.12.5.4 for information on setting up to read an external barometer).
Note
For differential mode operation (see Section 3.10.4.1) the AtmP selection
must be <4COM2>.
Summary of Contents for PG9000 Series
Page 10: ...PG9602 OPERATION AND MAINTENANCE MANUAL 2011 Fluke Calibration Page X Notes...
Page 128: ...PG9602 OPERATION AND MAINTENANCE MANUAL 2011 Fluke Calibration Page 118 Notes...
Page 164: ...PG9602 OPERATION AND MAINTENANCE MANUAL 2011 Fluke Calibration Page 154 Notes...
Page 188: ...PG9602 OPERATION AND MAINTENANCE MANUAL 2011 Fluke Calibration Page 178 Notes...
Page 192: ...PG9602 OPERATION AND MAINTENANCE MANUAL 2011 Fluke Calibration Page 182 Notes...