The main masses are loaded onto the mass loading tray in the opposite order
that they are loaded onto the PG9000. To install the masses on the mass
loading tray, proceed as follows:
Open the shipping cases.
Install the main masses:
The
main
masses (a series of 10 kg or 5 kg
masses) are vertically stacked on the mass loading spindle in reverse order
with the largest sequence number on the bottom of the stack and sequence
number 1 at the top.
The
main
masses are sequentially numbered to assist with identifying the
masses that are nominally the same. The sequence begins with number 1
and ends with number 8 (for a 100 kg MS). The
main
mass with the largest
sequential number is installed first at the bottom of the stack. The rest of the
main
masses should be stacked
upwards
in
descending
order ending with
main
mass sequential number
1
. Be careful not to confuse the M
ake Up
mass (refer to next item) with
main
mass
1
.
Install the other mass:
The
Make Up
mass is a single mass sequentially
numbered
1
. It has the same diameter as the
main
masses but is thinner. It
is a nominal 9 kg mass. A 55 kg or smaller mass set uses a 4 kg
make up
mass. The 9 kg
Make Up
mass is placed on top of the
main
mass stack. It is
always the top of the stack.
The idea of the make up mass is that when a mass load of 10 kg or greater is
required, the value of the make up mass plus the mass of the piston and bell
equal a nominal 10 kg. From an operational position this simplifies the mass
accounting responsibilities. From a visual perspective, the makeup mass
looks similar to the main masses and allows the operator to look at the mass
load and quickly sum the total nominal load.
Figure 7.
Mass sizes
Install the fractionary masses:
The
fractionary
masses are all the masses
of lower value than the 10 kg
main
masses and the 9 kg
Make Up
mass.
These are masses of 5 kg and under.
Fractionary
masses of 1 to 5 kg are
discs with a central hole.
Fractionary
masses of 100 to 500 g are solid, small
diameter pucks.
Fractionary
masses of 50 g and under are packed and
stored in their own separate storage case, they are commonly referred to as
trim masses.
Fractionary
mass discs and pucks are installed vertically in the corresponding slots
in the mass loading tray. Use a consistent setup for the sequence number
when there are two masses (e.g., always load sequential number
1
in the front).
Summary of Contents for PG9000 Series
Page 10: ...PG9602 OPERATION AND MAINTENANCE MANUAL 2011 Fluke Calibration Page X Notes...
Page 128: ...PG9602 OPERATION AND MAINTENANCE MANUAL 2011 Fluke Calibration Page 118 Notes...
Page 164: ...PG9602 OPERATION AND MAINTENANCE MANUAL 2011 Fluke Calibration Page 154 Notes...
Page 188: ...PG9602 OPERATION AND MAINTENANCE MANUAL 2011 Fluke Calibration Page 178 Notes...
Page 192: ...PG9602 OPERATION AND MAINTENANCE MANUAL 2011 Fluke Calibration Page 182 Notes...