Fluke P5500 Series User Manual Download Page 13

 

Hydraulic Comparison Test Pump 

 

Operation 

1

3

2

4

 

gjq002.eps 

Figure 2. Test Port Insert 

Table 3. Test Port Insert - Parts List 

Item Description 

P5515 

P5515-V P5515-EP 

1 O 

Ring 

3865142 

3865163 
 

3867888 

Test Port Insert 

3919892 

— 

— 

3 O 

Ring 

3883397 

3883521  3867895 

4 Test 

Port 

3918298 

— 

— 

To calibrate panel

-

mounted gauges with pressure connections in the rear, use an Angle 

Adapter (see Ancillary Equipment section). 

Operation 

Priming 

1.

 

Open reservoir valve (13) one turn counter

-

clockwise and turn screw press fully in. 

2.

 

Close valve (13) and turn screw press fully out. 

3.

 

Open valve and turn screw press fully in. 

Note 

During this operation, bubbles may appear in the reservoir, as trapped air 
is expelled. For large volumes, repeat steps 2 and 3 until no further bubbles 
appear. 

4.

 

With valve open, turn screw press fully out and close valve. The test pump is now 

ready for use. 

W

 Warning 

Turning the capstan out with the reservoir valve closed will 
generate approximately 15 inHg / 0.5 bar vacuum. If the DUT is 
vacuum sensitive, leave valve (13) open during priming 
operation. 

Summary of Contents for P5500 Series

Page 1: ...uke Corporation All rights reserved Printed in USA Specifications are subject to change without notice All product names are trademarks of their respective companies P5514 P5514 V P5514 EP Hydraulic C...

Page 2: ...vice center to obtain return authorization information then send the product to that service center with a description of the difficulty postage and insurance prepaid FOB Destination Fluke assumes no...

Page 3: ...ressed Liquid 2 Personal Protective Equipment 2 Symbols Used in this Manual 2 Connections 3 Connect the Gauge 4 Connect the Test Port Insert 4 Operation 5 Priming 5 Operation 6 Maintenance and Servici...

Page 4: ...P5514 P5514 V P5514 EP Users Manual ii...

Page 5: ...iii List of Tables Table Title Page 1 Operating Fluid Compatibility 2 2 Symbols 3 3 Test Port Insert Parts List 5 4 Seal Kit Part Numbers 6 5 Screw Press Assembly Parts List 10...

Page 6: ...P5514 P5514 V P5514 EP Users Manual iv...

Page 7: ...v List of Figures Figure Title Page 1 Connecting the Gauge 4 2 Test Port Insert 5 3 Reservoir and Test Port Assemblies 9 4 Screw Press Assembly 9 5 Angle Adapter 11 6 Pointer Remover Punch 11...

Page 8: ...P5514 P5514 V P5514 EP Users Manual vi...

Page 9: ...acy is maintained The comparison test pump should be mounted securely to a stable workbench or similar surface Four mounting holes are provided in the pump stand for this purpose How to Contact Fluke...

Page 10: ...ns that are not related to any specific procedures and do not appear elsewhere in this publication These are recommended precautions that personnel must understand and apply during equipment operation...

Page 11: ...fluids and protect the system from particulates we recommend the use of a Liquid to Liquid Separator see Section 5 Ancillary Equipment W Warning DO NOT use Teflon PTFE tape on these connections as th...

Page 12: ...the position to face forward hold the gauge adapter and turn the instrument COUNTER CLOCKWISE so that it faces forward 4 Hold the instrument steady while turning the gauge adapter COUNTER CLOCKWISE un...

Page 13: ...Open reservoir valve 13 one turn counter clockwise and turn screw press fully in 2 Close valve 13 and turn screw press fully out 3 Open valve and turn screw press fully in Note During this operation b...

Page 14: ...ution Never release the system pressure by using the reservoir valve alone as sudden depressurization will shock the system which may cause damage to sensitive instruments attached to it Maintenance a...

Page 15: ...t in the barrel 11 Re tighten the barrel union nut W Caution If the lead screw assembly 24 shows signs of excessive wear then it is very likely that the associated components have worn also therefore...

Page 16: ...n the barrel is worn also If the bore is worn or scratched it will not seal correctly and may leak under pressure 11 Re assembly is the reverse of the above Ensure that all seals and sealing surfaces...

Page 17: ...Hydraulic Comparison Test Pump Maintenance and Servicing 9 20 22 23 24 25 25 26 26 21 16 17 19 18 gjq004 eps Figure 4 Screw Press Assembly...

Page 18: ...lve Body 3923785 7 Bonded Seal 3862349 3922096 3922043 8 Reservoir 3919463 9 Locknut 3918244 10 Reservoir Cover 3918350 11 Washer 3916458 12 Spring 3922786 13 Valve Stem Assembly 3885982 14 Screw 3909...

Page 19: ...ould be used The angle adapter fits directly onto the test station converting it through 90 degrees allowing the standard adapters to be used gjn022 bmp Figure 5 Angle Adapter Pointer Remover Punch P5...

Page 20: ...P5514 P5514 V P5514 EP Users Manual 12...

Reviews: