Maintenance and Specifications
Maintenance
11
11-5
Optical Connector Care
•
Periodically clean and inspect the tester’s optical
connectors as described in Chapter 2.
•
Cover the connectors with the dust plugs when no
cable is connected.
•
Periodically clean the dust plugs with a lint-free swab
or wipe and isopropyl alcohol.
Replacing Reference Test Cords and Launch Fibers
Choose replacement reference test cords and launch and
receive fibers that meet the following requirements:
•
Core and cladding size: match the fiber to be tested
•
Connector polish: UPC for reference test cords and
launch fibers
•
Reference test cord length: minimum 2 m; maximum
5 m
•
Launch and receive fiber length: 100 m minimum for
multimode; 130 m minimum for singlemode
To ensure optimum performance from your tester, get
replacement reference test cords and launch fibers from
Fluke Networks.
Replacing the Battery
Replace the lithium ion battery pack when its life becomes
noticeably shorter. The battery is normally good for up to
400 charge/discharge cycles.
/
Note
Dispose of the lithium ion battery pack in
accordance with local regulations.
An internal lithium battery maintains the tester’s clock
and other settings when you remove the battery pack.
This battery typically lasts about 5 years. When the battery
begins to fail, the tester will lose the current date and
time when you remove the battery pack. If the lithium
battery fails, send the tester to a Fluke Networks service
center for a replacement.
Summary of Contents for OF-500 OptiFiber
Page 12: ...OF 500 OptiFiber Technical Reference Handbook x ...
Page 18: ...OF 500 OptiFiber Technical Reference Handbook xvi ...
Page 27: ...Getting Acquainted Powering the Tester 1 1 9 ajt20f eps Figure 1 1 Battery Pack Features ...
Page 29: ...Getting Acquainted Verifying Operation 1 1 11 ajt56f eps Figure 1 2 Removing the Module ...
Page 46: ...OF 500 OptiFiber Certifying OTDR Technical Reference Handbook 1 28 ...
Page 136: ...OF 500 OptiFiber Certifying OTDR Technical Reference Handbook 4 6 ...
Page 192: ...OF 500 OptiFiber Certifying OTDR Technical Reference Handbook 6 48 ...
Page 254: ...OF 500 OptiFiber Certifying OTDR Technical Reference Handbook 11 36 ...
Page 256: ...OF 500 OptiFiber Certifying OTDR Technical Reference Handbook ...
Page 272: ...B 14 OF 500 OptiFiber Certifying OTDR Technical Reference Handbook ...
Page 274: ...OF 500 OptiFiber Certifying OTDR Technical Reference Handbook C 2 ...
Page 282: ...OF 500 OptiFiber Technical Reference Handbook 8 ...