OF-500 OptiFiber Certifying OTDR
Technical Reference Handbook
6-44
1
On the
Loss/Length
tab in Setup, set the main and
remote testers to
Smart Remote
mode.
2
Clean the tester's and source's OUTPUT connectors
and the reference test cord connectors; then make
the connections shown in Figure 6-15.
3
On both testers press
F
, select FindFiber; then
press
T
.
4
Try various connections to the patch panel with the
main tester’s INPUT fiber until the input fiber’s
Link
Status
shows
CONNECTED
.
5
Then try various connections with the main tester’s
OUTPUT fiber until the output fiber’s
Link Status
shows
CONNECTED
. Table 6-11 shows the main tester
results for Smart Remote mode.
Table 6-11. Main Tester Results for FindFiber Test (Smart Remote Mode)
Link Status
INPUT Fiber
OUTPUT Fiber
Description
OPEN
UNKNOWN
The main tester’s INPUT fiber path is not complete. The main tester cannot determine
the state of the OUTPUT fiber path.
CONNECTED
OPEN
The main tester’s INPUT fiber path is complete. The main tester’s OUTPUT fiber path is
not complete.
CONNECTED
CONNECTED
Both fiber paths are complete.
Summary of Contents for OF-500 OptiFiber
Page 12: ...OF 500 OptiFiber Technical Reference Handbook x ...
Page 18: ...OF 500 OptiFiber Technical Reference Handbook xvi ...
Page 27: ...Getting Acquainted Powering the Tester 1 1 9 ajt20f eps Figure 1 1 Battery Pack Features ...
Page 29: ...Getting Acquainted Verifying Operation 1 1 11 ajt56f eps Figure 1 2 Removing the Module ...
Page 46: ...OF 500 OptiFiber Certifying OTDR Technical Reference Handbook 1 28 ...
Page 136: ...OF 500 OptiFiber Certifying OTDR Technical Reference Handbook 4 6 ...
Page 192: ...OF 500 OptiFiber Certifying OTDR Technical Reference Handbook 6 48 ...
Page 254: ...OF 500 OptiFiber Certifying OTDR Technical Reference Handbook 11 36 ...
Page 256: ...OF 500 OptiFiber Certifying OTDR Technical Reference Handbook ...
Page 272: ...B 14 OF 500 OptiFiber Certifying OTDR Technical Reference Handbook ...
Page 274: ...OF 500 OptiFiber Certifying OTDR Technical Reference Handbook C 2 ...
Page 282: ...OF 500 OptiFiber Technical Reference Handbook 8 ...