Power Analyzer
Specifications
81
Emissions that exceed the levels required by CISPR 11 can occur when the equipment is
connected to a test object.
Korea (KCC) ...................................... Class A Equipment
(Industrial Broadcasting & Communication Equipment)
Class A: Equipment meets requirements for industrial electromagnetic wave equipment and
the seller or user
should take notice of it. This equipment is intended for use in business environments and not
to be used in homes.
USA (FCC) ........................................ 47 CFR 15 subpart B.
This product is considered an exempt device per clause 15.103.
Electrical Specifications
Modules
NORMA 6003 .............................. 3 V 3 Current
NO ........................... 3 V 3 C 1 Motor
NORMA 6004 ............................. 4 V 4 Current
NORMA 6004+ ........................... 4 V 4 C 1 Motor
Sample Rate .................................... 200 kHz
Data Update Rate ............................ 100 ms, 200 ms, 500 ms, 1 s
Voltage
Range .............................................. 10 V, 100 V, 1000 V
Crest Factor .....................................
CF ≤ 2
Overload Capacity ............................. 10 % overload
Input Impedance ..............................
2 MΩ / 10 pF (Typical)
Temperature Coefficient .................. 0.05 x (Spec)/K @
<18°C or >28°C
Bandwidth ........................................ 1000 V range: 500 kHz
100 V range: 200 kHz
10 V range: 100 kHz