General Specifications
Size (HxWxL)..................................... 298 mm x 215 mm x 96 mm
Weight ............................................... 3.5 kg
Display ............................................. 5.7" TFT LCD, 640x480 pixels
Maximum voltage between any Terminal
and Earth Ground…........................... 1000 V
Temperature
Operating Temperature ................ 0 °C to 40 °C (with power adapter)
-10 °C to + 50 °C (battery only)
Storage Temperature ................... -30 °C to 60 °C (Without battery)
-20 °C to 60 °C (With battery)
Operating Humidity ............................ Non-condensing (<10 °C)
≤90 % RH (10 °C
至
30 °C)
≤75 % RH (30 °C
至
40 °C)
≤45 % RH (40 °C
至
50 °C)
Altitude
Operating ..................................... 2000 m
Storage ........................................ 12 000 m
Ingress Protection .............................. IEC 60529: IP50 (terminals mated)
Battery
Type ............................................ Li-ion, BP 291, 10.8V/5000 mAh, 54 Wh IEC 62133, UN 38.3
Battery life .................................... 10 hours (Typical)
Safety
General ........................................ IEC 61010-1: Pollution Degree 2
Measurement ............................... IEC 61010-2-030: CAT IV 600/CAT III 1000 V
Electromagnetic Compatibility (EMC)
International ....................................... IEC 61326-1: Industrial Electromagnetic Environment;
IEC 61326-2-2
,
CISPR 11: Group 1, Class A
Group 1: Equipment has intentionally generated and/or uses conductively-coupled radio
frequency energy that is necessary for the internal function of the equipment itself.
Class A: Equipment is suitable for use in all establishments other than domestic and those
directly connected to a low-voltage power supply network that supplies buildings used for
domestic purposes. There may be potential difficulties in ensuring electromagnetic
compatibility in other environments due to conducted and radiated disturbances.
Caution: This equipment is not intended for use in residential environments and may not
provide adequate protection to radio reception in such environments.