Remote Control
Users Guide
4-36
OUTPut9[:STATe] ON | OFF
Description
Sets the state of the synchronization output. Can only be set to ON if SYNC:SOURce is
not set to EXTernal.
Parameter(s)
ON
The synchronization pulses are output at the rear sync input/output jack.
OFF
No synchronization pulses are output.
*RST state
OFF
Example(s)
OUTP9 ON
OUTP9? Response:
0
Invalidates
-
Invalidated by
SYNC:SOURce EXTernal
ROUTe Subsystem
The ROUTe subsystem selects the type of connection the instrument uses to measure on a
three phase system.
Command Parameter
Default
Value/Unit
Remark
:ROUTe
:SYSTem
“3W” | “2W”
“3W”
ROUTe:SYSTem “3W” | “2W”
Description
Selects the type of connection the instrument uses to measure on a three phase system.
Note
Parameter 2W is only accepted by firmware version 1.4 and higher.
Parameter(s)
“3W” Three-wattmeter
configuration.
“2W” Two-wattmeter
configuration.
*RST state
“3W”
Example(s)
ROUT:SYST “3W”
ROUT:SYST? Response:
“3W”