Remote Control
Users Guide
4-64
SYNC Subsystem
The SYNC subsystem controls the synchronization capability of the instrument. When
synchronization is enabled, the instrument adapts the averaging cycles to the frequency of
the signal fed to synchronization source. If memory recording of sampled data is in
operation, the synchronization signal can be used as a specific form of triggering.
Command Parameter
Default
Value/Unit
Remark
:SYNC
:STATe ON
|
OFF
ON
:LEVel
:UNIT ABSolute
|
PCT PCT
[:SOURce]|VOLTage[1..6]| VOLTage[1..6]|CURRent[1..6]| VOLTage1
SOURce = current
CURRent[1..6]|EXTernal EXTernal
source
<curr. only SOURce
node is implemented>
:LEVel
-150%...150% of range
0.0
Not for EXTernal
:SLOPe POSitive
|
NEGative
POSitive
:FILTer
Not for EXTernal source
[:LPASs]
[:STATe] ON
|
OFF
OFF
:FREQuency
1.0e2, 1.0e3, 1.0e4
1.0e4 Hz
:TIMeout
0.015 to 3600.0
0.3 s
SYNC:STATe ON | OFF
Description
Sets whether the averaging interval is controlled by signal frequency on selected input or
not. If synchronization is turned on, the actual averaging period is kept to be a first
integer multiple of the sync signal greater than user-specified nominal averaging period.
If synchronization is turned off, the actual averaging period is equal to the user-specified
nominal averaging period rounded to integer multiple of sample periods.
Parameter(s)
ON
Synchronization is required. The instrument will always attempt to synchronize to the
sync source signal frequency.
OFF
Synchronization is disabled. Use this option for measurements on DC signals.
Example(s)
SYNC:STAT ON
SYNC:STAT? Response:
1
*RST state
ON
Summary of Contents for NORMA 4000 Operators
Page 6: ...Remote Control Users Guide iv ...
Page 8: ...Remote Control Users Guide vi ...
Page 10: ...Remote Control Users Guide viii ...
Page 26: ...Remote Control Users Guide 2 2 ...
Page 38: ...Remote Control Users Guide 3 2 ...
Page 50: ...Remote Control Users Guide 4 2 ...
Page 150: ...Remote Control Users Guide 4 102 ...
Page 152: ...Remote Control Users Guide 5 2 ...
Page 160: ...Remote Control Users Guide 6 2 ...
Page 184: ...Remote Control Users Guide 6 26 ...