![Fluke NetDAQ 2640A Service Manual Download Page 29](http://html1.mh-extra.com/html/fluke/netdaq-2640a/netdaq-2640a_service-manual_2310217029.webp)
Introduction and Specification
Specifications
1
1-13
2640A Specifications
1-20.
This section includes specifications specific to the 2640A instrument by measurement
function.
2640A DC Voltage Measurement Specifications
1-21.
Tables 1-13 to 1-15 provide 2640A specifications for the dc voltage measurement
function.
Table 1-13. 2640A DC Voltage General Specifications
Specification
Characteristic
Input Impedance
100 M
Ω
in parallel with 300 pF maximum for ranges <=3V
10 M
Ω
in parallel with 100 pF maximum for ranges >3V
Normal Mode Rejection
50 dB minimum at 50 Hz/60 Hz +0.1%, Slow Rate
Common Mode Rejection
120 dB minimum at dc, 50 Hz/60 Hz +0.1%, 1 k
Ω
imbalance, Slow
Rate
80 dB minimum at dc, 50 Hz/60 Hz +0.1%, 1 k
Ω
imbalance, Medium
and Fast Rates
Channel-to-Channel Crosstalk
120 dB minimum Slow Rate (e.g., 30V dc on channel 1 may cause a
30
µ
V error on channel 2)
100 dB minimum Medium and Fast Rates (e.g., 1V dc on channel 1
may cause a 10
µ
V error on channel 2)
Temperature Coefficient
For % input: Add 1/10th the 90-day specification per
°
C above 28
°
C
or below 18
°
C
For floor error (V): Add 1/20th the 90-day specification per
°
C above
28
°
C or below 18
°
C
Maximum Input Voltage
The lesser voltage of:
300V from any terminal on channels 1 and 11 to earth;
150V from any terminal on channels 2 through 10, and 12 through 20
to earth;
300V from any terminal on channels 1 and 11 to any other terminal;
150V from any terminal on channels 2 through 10, and 12 through 20
to any other input terminal
Table 1-14. 2640A DC Voltage Range and Resolution Specifications
Resolution
Range
Slow
Fast
90 mV
0.3
µ
V
1
µ
V
300 mV
1
µ
V
3
µ
V
3V
10
µ
V
30
µ
V
30V
100
µ
V
300
µ
V
150V/300V
1 mV
3 mV
Note 300V range applies to channels 1 and 11 only.
Summary of Contents for NetDAQ 2640A
Page 14: ...NetDAQ Service Manual x...
Page 46: ...NetDAQ Service Manual 2 4...
Page 106: ...NetDAQ Service Manual 2 64...
Page 108: ...NetDAQ Service Manual 3 2...
Page 164: ...NetDAQ Service Manual 4 42...
Page 206: ...NetDAQ Service Manual 6 2...
Page 218: ...NetDAQ Service Manual 6 14 2645A 1601 Figure 6 2 A1 Main PCA Assembly...
Page 220: ...NetDAQ Service Manual 6 16 2620A 1601 Figure 6 3 A2 Display PCA Assembly...
Page 230: ...NetDAQ Service Manual 6 26 2645A 1603 Figure 6 5 2645A A3 A D Converter PCA Assembly...
Page 234: ...NetDAQ Service Manual 7 2...
Page 242: ...2640A 2645A Service Manual 7 10 Figure 7 2 A2 Display PCA Assembly 2620A 1602...
Page 243: ...Schematic Diagrams 7 7 11 Figure 7 2 A2 Display PCA Assembly cont 2620A 1002...
Page 251: ...Schematic Diagrams 7 7 19 2645A 1603 Figure 7 4 2645A A3 A D Converter PCA Assembly 2645A 1603...
Page 258: ...2640A 2645A Service Manual 7 26 Figure 7 5 A4 Analog Input PCA Assembly 2620A 1604...
Page 260: ...2640A 2645A Service Manual 7 28...