1
1-9
2640A/2645A General Specifications
1-11.
Table 1-5 provides the general specifications for the 2640A and 2645A instruments.
Table 1-5. 2640A/2645A General Specifications
Specification
Characteristic
Channel Capacity
20
I/O Lines Total
12
Size
9.3 cm (3.67 in) high, 21.6 cm (8.5 in) wide, 36.2 cm (14.28 in) deep
Weight
Net, 4 kg (8.8 lb.) Shipping, 6.0 kg (13.2 lb.)
Power
107 to 264V ac (no switching required), 45 to 65 Hz, 15 VA maximum
9V dc to 16V dc, 6W maximum. Specifications are for 50 or 60 Hz operation.
If both sources are applied simultaneously, ac voltage is used if it exceeds
approximately 8 times the dc voltage. Automatic switchover occurs between ac
and dc without interruption.
Standards
Both instruments comply with:
IEC 1010-1
UL 1244
CSA Bulletin 556B
ANSI/ISA-S8201-1988
CSA C22.2 No. 101.1-92
Vfg. 243/1991 (when shielded cables are used)
FCC-15B, Class B level (when shielded cables are used)
Serial Interface
(RS-232C)
Connector: 9 pin male (DB-9P)
Signals: TX, RX, DTR, RTS, GND
Modem Control: full duplex
Baud rates: 4800, 9600, 19200, 38400
Data format: 8 data bits, no parity bit, one stop bit
Flow control: XON/XOFF
Echo: Off
Common Mode
Voltage
2640A 150V (300V on channels 1 and 11)
2645A 50V dc or 30V ac rms.
Measurement Speed
(Scanning Rates)
2640A
Slow - 6 readings per second
Medium - 48 readings per second (60 Hz)
Fast - 143 readings per second (20 configured channels)
Single Channel - 120 readings per second
2645A
Slow - 54 readings per second (60 Hz)
Medium - 200 readings per second
Fast - 1000 readings per second (20 configured channels)
Single Channel - 400 readings per second
Accuracy of Medium
Scanning Rate
Equal to (Fast Accuracy Rate + Slow Accuracy Rate)/2
Additional error if
“Automatic drift
correction” is turned
off.
If the instrument is fully warmed-up at the time drift correction was disabled, i.e.,
turned on at least 1 hour earlier: 1/10 of the 90-day specification per
°
C change
in ambient temperature from the temperature when drift correction was disabled.
If the instrument was not fully warmed up at the time of drift correction was
disabled: add an error equal to the 90-day specification for instrument warmup
+1/10 of the 90-day specification per
°
C change in ambient temperature from the
temperature when drift correction was disabled.
Summary of Contents for NetDAQ 2640A
Page 14: ...NetDAQ Service Manual x...
Page 46: ...NetDAQ Service Manual 2 4...
Page 106: ...NetDAQ Service Manual 2 64...
Page 108: ...NetDAQ Service Manual 3 2...
Page 164: ...NetDAQ Service Manual 4 42...
Page 206: ...NetDAQ Service Manual 6 2...
Page 218: ...NetDAQ Service Manual 6 14 2645A 1601 Figure 6 2 A1 Main PCA Assembly...
Page 220: ...NetDAQ Service Manual 6 16 2620A 1601 Figure 6 3 A2 Display PCA Assembly...
Page 230: ...NetDAQ Service Manual 6 26 2645A 1603 Figure 6 5 2645A A3 A D Converter PCA Assembly...
Page 234: ...NetDAQ Service Manual 7 2...
Page 242: ...2640A 2645A Service Manual 7 10 Figure 7 2 A2 Display PCA Assembly 2620A 1602...
Page 243: ...Schematic Diagrams 7 7 11 Figure 7 2 A2 Display PCA Assembly cont 2620A 1002...
Page 251: ...Schematic Diagrams 7 7 19 2645A 1603 Figure 7 4 2645A A3 A D Converter PCA Assembly 2645A 1603...
Page 258: ...2640A 2645A Service Manual 7 26 Figure 7 5 A4 Analog Input PCA Assembly 2620A 1604...